dc.description.abstractPresented study describes the optimization method of an HPGe detector through implementation of Geant4 toolkit. The optimized model was verified through comparison with experimentally obtained data using a set of point-like radioactive calibration sources. Acquired results displayed good agreement with the experimental data that falls under an average relative deviation of the order of ~ 2% within the energy range of 53–1836 keV. Additionally, in order to test the validity of the code it was also applied to a different detection equipment where an average relative deviation of the order of ~ 1.8% was achieved within the energy range of 121–1112 keV. © 2023, Akadémiai Kiadó, Budapest, Hungary.en
dc.language.isoen
dc.relationinfo:eu-repo/grantAgreement/MESTD/Basic Research (BR or ON)/171002/RS//
dc.relationIAEA Coordinated Research Project [RC 23159]
dc.rightsrestrictedAccess
dc.sourceJournal of Radioanalytical and Nuclear Chemistry
dc.subjectFEP efficiencyen
dc.subjectGeant4 modelen
dc.subjectHPGe detectoren
dc.subjectMonte Carlo simulationen
dc.subjectγ-Spectrometryen
dc.titleDetailed optimization procedure of an HPGe detector using Geant4 toolkiten
dc.typearticleen
dc.rights.licenseARR
dc.citation.volume332
dc.citation.issue3
dc.citation.spage817
dc.citation.epage828
dc.identifier.wos000943845400001
dc.identifier.doi10.1007/s10967-023-08810-x
dc.type.versionpublishedVersion
dc.identifier.scopus2-s2.0-85149310591


Документи

Thumbnail

Овај документ се појављује у следећим колекцијама

Приказ основних података о документу