Detailed optimization procedure of an HPGe detector using Geant4 toolkit
Samo za registrovane korisnike
2023
Autori
Travar, M.Nikolov, Jovana
Todorović, N.
Vraničar, A.
Völgyesi, P.
Kirchknopf, P.
Čeliković, Igor T.
Milanović, Tamara J.
Joković, D.
Članak u časopisu (Objavljena verzija)
Metapodaci
Prikaz svih podataka o dokumentuApstrakt
Presented study describes the optimization method of an HPGe detector through implementation of Geant4 toolkit. The optimized model was verified through comparison with experimentally obtained data using a set of point-like radioactive calibration sources. Acquired results displayed good agreement with the experimental data that falls under an average relative deviation of the order of ~ 2% within the energy range of 53–1836 keV. Additionally, in order to test the validity of the code it was also applied to a different detection equipment where an average relative deviation of the order of ~ 1.8% was achieved within the energy range of 121–1112 keV. © 2023, Akadémiai Kiadó, Budapest, Hungary.
Ključne reči:
FEP efficiency / Geant4 model / HPGe detector / Monte Carlo simulation / γ-SpectrometryIzvor:
Journal of Radioanalytical and Nuclear Chemistry, 2023, 332, 3, 817-828Finansiranje / projekti:
- Nuklearne metode istraživanja retkih događaja i kosmičkog zračenja (RS-MESTD-Basic Research (BR or ON)-171002)
- IAEA Coordinated Research Project [RC 23159]
DOI: 10.1007/s10967-023-08810-x
ISSN: 0236-5731
WoS: 000943845400001
Scopus: 2-s2.0-85149310591
Institucija/grupa
VinčaTY - JOUR AU - Travar, M. AU - Nikolov, Jovana AU - Todorović, N. AU - Vraničar, A. AU - Völgyesi, P. AU - Kirchknopf, P. AU - Čeliković, Igor T. AU - Milanović, Tamara J. AU - Joković, D. PY - 2023 UR - https://vinar.vin.bg.ac.rs/handle/123456789/10734 AB - Presented study describes the optimization method of an HPGe detector through implementation of Geant4 toolkit. The optimized model was verified through comparison with experimentally obtained data using a set of point-like radioactive calibration sources. Acquired results displayed good agreement with the experimental data that falls under an average relative deviation of the order of ~ 2% within the energy range of 53–1836 keV. Additionally, in order to test the validity of the code it was also applied to a different detection equipment where an average relative deviation of the order of ~ 1.8% was achieved within the energy range of 121–1112 keV. © 2023, Akadémiai Kiadó, Budapest, Hungary. T2 - Journal of Radioanalytical and Nuclear Chemistry T1 - Detailed optimization procedure of an HPGe detector using Geant4 toolkit VL - 332 IS - 3 SP - 817 EP - 828 DO - 10.1007/s10967-023-08810-x ER -
@article{ author = "Travar, M. and Nikolov, Jovana and Todorović, N. and Vraničar, A. and Völgyesi, P. and Kirchknopf, P. and Čeliković, Igor T. and Milanović, Tamara J. and Joković, D.", year = "2023", abstract = "Presented study describes the optimization method of an HPGe detector through implementation of Geant4 toolkit. The optimized model was verified through comparison with experimentally obtained data using a set of point-like radioactive calibration sources. Acquired results displayed good agreement with the experimental data that falls under an average relative deviation of the order of ~ 2% within the energy range of 53–1836 keV. Additionally, in order to test the validity of the code it was also applied to a different detection equipment where an average relative deviation of the order of ~ 1.8% was achieved within the energy range of 121–1112 keV. © 2023, Akadémiai Kiadó, Budapest, Hungary.", journal = "Journal of Radioanalytical and Nuclear Chemistry", title = "Detailed optimization procedure of an HPGe detector using Geant4 toolkit", volume = "332", number = "3", pages = "817-828", doi = "10.1007/s10967-023-08810-x" }
Travar, M., Nikolov, J., Todorović, N., Vraničar, A., Völgyesi, P., Kirchknopf, P., Čeliković, I. T., Milanović, T. J.,& Joković, D.. (2023). Detailed optimization procedure of an HPGe detector using Geant4 toolkit. in Journal of Radioanalytical and Nuclear Chemistry, 332(3), 817-828. https://doi.org/10.1007/s10967-023-08810-x
Travar M, Nikolov J, Todorović N, Vraničar A, Völgyesi P, Kirchknopf P, Čeliković IT, Milanović TJ, Joković D. Detailed optimization procedure of an HPGe detector using Geant4 toolkit. in Journal of Radioanalytical and Nuclear Chemistry. 2023;332(3):817-828. doi:10.1007/s10967-023-08810-x .
Travar, M., Nikolov, Jovana, Todorović, N., Vraničar, A., Völgyesi, P., Kirchknopf, P., Čeliković, Igor T., Milanović, Tamara J., Joković, D., "Detailed optimization procedure of an HPGe detector using Geant4 toolkit" in Journal of Radioanalytical and Nuclear Chemistry, 332, no. 3 (2023):817-828, https://doi.org/10.1007/s10967-023-08810-x . .