@conference{
author = "Haralambos Apostolakopoulos, Filip and Kržanović, Nikola and Perazić, Luka and Živanović, Miloš and Stanković, Koviljka",
year = "2016",
abstract = "Determination of the response of thermoluminescent and electronic personal dosimeters as a function of photon energy and angle of incidence has a crucial impact on their application in poly-energetic multidirectional photon radiation fields. In the experimental part of this paper, in order to determine the energy dependence of the dosimeter responses, the dosimeters were mounted on an ISO slab phantom, irradiated in X and gamma ray fields, defined in the IAEA SRS 16, within the energy range of 33 keV - 1,25 MeV. Tested dosimeters have not shown an adequate response in the low photon energies relative to their Cs-137 responses, while in the middle and high energy photon ranges they performed well and have shown a deviation relative to their Cs-137 responses up to ±20%. In order to determine the angular dependence of the dosimeters the N-200 narrow beam radiation quality has been used. The angles of incidence for which the dosimeters were tested range from 0° to 80°, with an increment of 20°. Most of the dosimeters have behaved as expected, except for certain dosimeters due to their significant increase in response caused by the influence of the scattered radiation.",
publisher = "Belgrade : ETRAN Society",
journal = "IcETRAN 2016 : 3rd International Conference on Electrical, Electronic and Computing Engineering; Proceedings book",
title = "Comparison of the Energy and Angular Responses of Thermoluminescent and Electronic Personal Dosimeters",
url = "https://hdl.handle.net/21.15107/rcub_vinar_11915"
}