Приказ основних података о документу

dc.creatorStepanić, Nenad
dc.creatorTerzić, Marijana M.
dc.creatorRadivojević, Dušan
dc.creatorRaković, Dejan
dc.date.accessioned2018-12-10T12:13:18Z
dc.date.available2018-12-10T12:13:18Z
dc.date.issued2018
dc.identifier.issn0354-9836
dc.identifier.urihttp://www.doiserbia.nb.rs/Article.aspx?ID=0354-98361800176S
dc.identifier.urihttps://vinar.vin.bg.ac.rs/handle/123456789/7972
dc.description.abstractThis paper presents design, development, and testing of a custom-made calibration environment for the calibration of temperature sensors applied in a guarded hot plate apparatus for thermal conductivity measurements. Description of an in situ calibration principle and realization are introduced. Results of numerical simulation, as well as those of experimental validation are given in a separate section of the paper. According to the results, proposed in situ calibration by using the applied guarded hot plate apparatus elements improves both accuracy and traceability the of thermal conductivity measurement. © 2018 Serbian Society of Heat Transfer Engineers.en
dc.rightsopenAccess
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/4.0/
dc.sourceThermal Science
dc.subjectguarded hot plate methoden
dc.subjectin situ temperature calibrationen
dc.titleAn in situ temperature calibration of a guarded hot plate apparatusen
dc.typearticleen
dc.rights.licenseBY-NC-ND
dcterms.abstractРаковиц, Дејан; Терзиц, Маријана; Степаниц, Ненад; Радивојевиц, Дусан;
dc.rights.holder© 2018 Serbian Society of Heat Transfer Engineers
dc.citation.volume22
dc.citation.issue00 (Online first)
dc.citation.spage176
dc.citation.epage176
dc.identifier.wos000462414100025
dc.identifier.doi10.2298/TSCI180425176S
dc.citation.rankM22
dc.type.versionpublishedVersion
dc.identifier.scopus2-s2.0-85057122654
dc.identifier.fulltexthttps://vinar.vin.bg.ac.rs//bitstream/id/10842/0354-98361800176S.pdf