Influence of anode/filtration setup on X-ray multimeter energy response in mammography applications
2024
Преузимање 🢃
Аутори
Kojić, Andrea
Kržanović, Nikola
Živanović, Miloš
Toroi, Paula
Bakrač, Luka
Božović, Predrag
Stanković Petrović, Jelena
Чланак у часопису (Објављена верзија)
Метаподаци
Приказ свих података о документуАпстракт
Quality control and assurance of mammography X-ray generators include usage of solid-state detectors and/or ionization chambers which are calibrated in standard reference radiation fields. IEC 61267:2005 standard defines reference mammography radiation fields for Mo/Mo anode/filtration, while various anode/filtration combinations are encountered in clinical mammography X-ray units. Not all Secondary Standard Dosimetry Laboratories have an X-ray generator with Mo/Mo radiation setup, thus traceability can be established only for the available anode/filter combinations which is commonly limited to W/Al. In this study, W/Al radiation fields were established under laboratory conditions by performing half-value layer measurements, and characterization of four solid-state detectors was performed. X-ray multimeter performance was evaluated in terms of energy response in the W/Al radiation fields. In the laboratory conditions the energy response of the dosimeters had larger deviation from unity... for dosimeters without appropriate anode/filter combination selected in the software settings, even though most of the dosimeters had uniform relative response. Discrepancy in the response was further investigated by examining its variation induced by available detector software settings, and it was determined that the dosimeter response can vary up to 20 %. In clinical setup, half-value layer was determined, and detector performance was examined. Dosimeters were tested in clinical fields with Mo/Mo, Mo/Rh, W/Rh, W/Ag anode/filter combinations in the X-ray tube voltage range from 25 kV to 35 kV. For most clinical radiation fields, multi-element detectors had energy response deviation within ±5 %. The single-element detector had one software setting available and has exhibited strong energy dependence. Extensive testing of detector response such as presented in this paper allows for correction factor interpolation based on half-value layer.
Кључне речи:
anode/filter setup / detector response / mammography / solid-state detectorИзвор:
Radiation Measurements, 2024, 174, 107135-Финансирање / пројекти:
- Министарство науке, технолошког развоја и иновација Републике Србије, институционално финансирање - 200017 (Универзитет у Београду, Институт за нуклеарне науке Винча, Београд-Винча) (RS-MESTD-inst-2020-200017)
- International Atomic Energy Agency (IAEA) [contract No: 24688]
- European Partnership on Metrology [project 22NRM01 TraMeXi]
Напомена:
Повезане информације:
- Повезани садржај
https://vinar.vin.bg.ac.rs/handle/123456789/13372
DOI: 10.1016/j.radmeas.2024.107135
ISSN: 1350-4487
WoS: 001235048700001
Scopus: 2-s2.0-85191324332
Колекције
Институција/група
VinčaTY - JOUR AU - Kojić, Andrea AU - Kržanović, Nikola AU - Živanović, Miloš AU - Toroi, Paula AU - Bakrač, Luka AU - Božović, Predrag AU - Stanković Petrović, Jelena PY - 2024 UR - https://vinar.vin.bg.ac.rs/handle/123456789/13192 AB - Quality control and assurance of mammography X-ray generators include usage of solid-state detectors and/or ionization chambers which are calibrated in standard reference radiation fields. IEC 61267:2005 standard defines reference mammography radiation fields for Mo/Mo anode/filtration, while various anode/filtration combinations are encountered in clinical mammography X-ray units. Not all Secondary Standard Dosimetry Laboratories have an X-ray generator with Mo/Mo radiation setup, thus traceability can be established only for the available anode/filter combinations which is commonly limited to W/Al. In this study, W/Al radiation fields were established under laboratory conditions by performing half-value layer measurements, and characterization of four solid-state detectors was performed. X-ray multimeter performance was evaluated in terms of energy response in the W/Al radiation fields. In the laboratory conditions the energy response of the dosimeters had larger deviation from unity for dosimeters without appropriate anode/filter combination selected in the software settings, even though most of the dosimeters had uniform relative response. Discrepancy in the response was further investigated by examining its variation induced by available detector software settings, and it was determined that the dosimeter response can vary up to 20 %. In clinical setup, half-value layer was determined, and detector performance was examined. Dosimeters were tested in clinical fields with Mo/Mo, Mo/Rh, W/Rh, W/Ag anode/filter combinations in the X-ray tube voltage range from 25 kV to 35 kV. For most clinical radiation fields, multi-element detectors had energy response deviation within ±5 %. The single-element detector had one software setting available and has exhibited strong energy dependence. Extensive testing of detector response such as presented in this paper allows for correction factor interpolation based on half-value layer. T2 - Radiation Measurements T1 - Influence of anode/filtration setup on X-ray multimeter energy response in mammography applications VL - 174 SP - 107135 DO - 10.1016/j.radmeas.2024.107135 ER -
@article{
author = "Kojić, Andrea and Kržanović, Nikola and Živanović, Miloš and Toroi, Paula and Bakrač, Luka and Božović, Predrag and Stanković Petrović, Jelena",
year = "2024",
abstract = "Quality control and assurance of mammography X-ray generators include usage of solid-state detectors and/or ionization chambers which are calibrated in standard reference radiation fields. IEC 61267:2005 standard defines reference mammography radiation fields for Mo/Mo anode/filtration, while various anode/filtration combinations are encountered in clinical mammography X-ray units. Not all Secondary Standard Dosimetry Laboratories have an X-ray generator with Mo/Mo radiation setup, thus traceability can be established only for the available anode/filter combinations which is commonly limited to W/Al. In this study, W/Al radiation fields were established under laboratory conditions by performing half-value layer measurements, and characterization of four solid-state detectors was performed. X-ray multimeter performance was evaluated in terms of energy response in the W/Al radiation fields. In the laboratory conditions the energy response of the dosimeters had larger deviation from unity for dosimeters without appropriate anode/filter combination selected in the software settings, even though most of the dosimeters had uniform relative response. Discrepancy in the response was further investigated by examining its variation induced by available detector software settings, and it was determined that the dosimeter response can vary up to 20 %. In clinical setup, half-value layer was determined, and detector performance was examined. Dosimeters were tested in clinical fields with Mo/Mo, Mo/Rh, W/Rh, W/Ag anode/filter combinations in the X-ray tube voltage range from 25 kV to 35 kV. For most clinical radiation fields, multi-element detectors had energy response deviation within ±5 %. The single-element detector had one software setting available and has exhibited strong energy dependence. Extensive testing of detector response such as presented in this paper allows for correction factor interpolation based on half-value layer.",
journal = "Radiation Measurements",
title = "Influence of anode/filtration setup on X-ray multimeter energy response in mammography applications",
volume = "174",
pages = "107135",
doi = "10.1016/j.radmeas.2024.107135"
}
Kojić, A., Kržanović, N., Živanović, M., Toroi, P., Bakrač, L., Božović, P.,& Stanković Petrović, J.. (2024). Influence of anode/filtration setup on X-ray multimeter energy response in mammography applications. in Radiation Measurements, 174, 107135. https://doi.org/10.1016/j.radmeas.2024.107135
Kojić A, Kržanović N, Živanović M, Toroi P, Bakrač L, Božović P, Stanković Petrović J. Influence of anode/filtration setup on X-ray multimeter energy response in mammography applications. in Radiation Measurements. 2024;174:107135. doi:10.1016/j.radmeas.2024.107135 .
Kojić, Andrea, Kržanović, Nikola, Živanović, Miloš, Toroi, Paula, Bakrač, Luka, Božović, Predrag, Stanković Petrović, Jelena, "Influence of anode/filtration setup on X-ray multimeter energy response in mammography applications" in Radiation Measurements, 174 (2024):107135, https://doi.org/10.1016/j.radmeas.2024.107135 . .