Evaluacija performansi monitora zračenja prema pravilniku o dozimetrima
Survey meter performance evaluation according to the national regulation on radiation protection dosimeters
| dc.creator | Kržanović, Nikola | |
| dc.creator | Živanović, Miloš | |
| dc.creator | Đaletić, Miloš | |
| dc.creator | Komatina, Ivana | |
| dc.creator | Stanković, Srboljub | |
| dc.creator | Vlahović, Jelena | |
| dc.date.accessioned | 2025-12-09T12:55:42Z | |
| dc.date.available | 2025-12-09T12:55:42Z | |
| dc.date.issued | 2025 | |
| dc.identifier.isbn | 978-86-7306-178-8 | |
| dc.identifier.uri | https://vinar.vin.bg.ac.rs/handle/123456789/15962 | |
| dc.description.abstract | Pravilnikom o dozimetrima je u oblast zaštite od zračenja uvedena zakonska regulativa za aktivne elektronske dozimetre, u smislu njihovog periodičnog etaloniranja i overavanja. Takođe, Pravilnikom je poboljšana dozimetrijska praksa u oblasti zaštite od zračenja usled postavljanja kriterijuma prihvatljivosti za aktivne elektronske dozimetre u vidu najveće dozvoljene greške. Izvršeno je ispitivanje performansi pet najčešće korišćenih modela monitora zračenja u širokom opsegu jačina ambijentalnog ekvivalenta doze, kao i u širokom opsegu energija od N-40 (33,3 keV) do S-Co (1,25 MeV). Ispitivani dozimetri su overeni, pošto su pokazali zadovoljavajuće dozimetrijske karakteristike. | sr |
| dc.description.abstract | Through the national regulation on radiation protection dosimeters, a legal framework was established, enforcing their periodic calibration and verification. By introducing the verification acceptability criteria, defined by the national regulation document, and expressed in terms of maximum permissible error, the radiation protection dosimetry practice was improved. Performance assessment of five most commonly used survey meters was done in a wide range of ambient dose equivalent rate and a wide photon energy range from N-40 (33.3 keV) to S-Co (1.25 MeV). Examined dosimeters were successfully verified, displaying satisfactory dosimetric properties. | en |
| dc.language.iso | en | sr |
| dc.publisher | Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе | sr |
| dc.rights | openAccess | sr |
| dc.rights.uri | https://creativecommons.org/licenses/by-nc-nd/4.0/ | |
| dc.source | XXXIII симпозијум ДЗЗСЦГ : зборник радова | sr |
| dc.title | Evaluacija performansi monitora zračenja prema pravilniku o dozimetrima | sr |
| dc.title | Survey meter performance evaluation according to the national regulation on radiation protection dosimeters | en |
| dc.type | conferenceObject | sr |
| dc.rights.license | BY-NC-ND | sr |
| dc.rights.holder | © Institut za nuklearne nauke „Vinča“ - Institut od nacionalnog značaja za Republiku Srbiju, Univerzitet u Beogradu | sr |
| dc.citation.spage | 345 | |
| dc.citation.epage | 351 | |
| dc.description.other | XXXIII симпозијум ДЗЗСЦГ [Друштва за заштиту од зрачења Србије и Црне Горе] : 1-3. октобар 2025. године. | sr |
| dc.type.version | publishedVersion | sr |
| dc.identifier.fulltext | http://vinar.vin.bg.ac.rs/bitstream/id/45078/DZZSCG-25.pdf | |
| dc.identifier.rcub | https://hdl.handle.net/21.15107/rcub_vinar_15962 |
Files in this item
This item appears in the following Collection(s)
-
Zbornici Društva za zaštitu od zračenja Srbije i Crne Gore
DZZ SCG -
Radovi istraživača
Researchers' publications