Show simple item record

dc.creatorTomljenovic, II
dc.creatorNinković, Marko M.
dc.creatorBek-Uzarov, D
dc.creatorStanković, Srboljub J.
dc.creatorKovačević, Milojko
dc.date.accessioned2018-03-01T18:44:56Z
dc.date.available2018-03-01T18:44:56Z
dc.date.issued1999
dc.identifier.issn0031-9155
dc.identifier.urihttps://vinar.vin.bg.ac.rs/handle/123456789/2276
dc.description.abstractExperimentally determined values of x-ray backscatter factors are presented in this paper. Measurements were made for x-rays generated at voltages between 60 kV and 300 kV and three water phantoms different in shape and size. To study the influence of the irradiation geometry on the backscatter factors, the measurements were performed for different photon beam held diameters at the phantom front face, at a fixed source-to-phantom distance of 1 m. An ionization chamber of volume 1 cm(3) was used for dose measurement. Measured results are analysed and discussed in comparison with measured and calculated values given in the cited references.en
dc.rightsrestrictedAccessen
dc.sourcePhysics in Medicine and Biologyen
dc.titleWater phantom backscatter factors for x-rays in the 60 kV to 300 kV regionen
dc.typearticleen
dcterms.abstractНинковиц, ММ; Станковиц, СЈ; Ковацевиц, М; Томљеновиц, ИИ; Бек-Узаров, Д;
dc.citation.volume44
dc.citation.issue9
dc.citation.spage2193
dc.citation.epage2200
dc.identifier.wos000082508100007
dc.identifier.doi10.1088/0031-9155/44/9/307
dc.citation.rankM21
dc.identifier.pmid10495114
dc.identifier.scopus2-s2.0-0032878123