Screen printed diamond electrode as efficient “point-of-care” platform for submicromolar determination of cytostatic drug in biological fluids and pharmaceutical product

Само за регистроване кориснике
2021
Аутори
Stanković, Dalibor M.
Milanović, Zorana
Švorc, Lubomir
Stanković, Vesna
Janković, Drina
Mirković, Marija D.
Vranješ-Đurić, Sanja
Чланак у часопису (Објављена верзија)
,
© 2021 Elsevier B.V.
Метаподаци
Приказ свих података о документу
Апстракт
This work presents advanced electrochemical platform based on screen printed diamond electrode (SPDE) system for the single drop “point-of-care” testing. Proposed approach was applied for the quantification of doxorubicin hydrochloride (DOX) in biological fluids and pharmaceutical product. Using a single drop (~30 μL) of the tested solution at the electrode surface, DOX showed high electroactivity over a wide range of pHs. In these conditions, single oval shaped, well-defined and pH dependent oxidation peak was observed in the potential range from 0.5 V to 1.3 V. In the reverse scan, two cathodic peaks, were noted – around 0.3 V and – 0.5 V. Similarly, first reduction peak was pH dependent, while second one was independent in the studied range. Experimental conditions for DOX quantification were optimized and natures of the electrode reactions were investigated. Working linear range obtained for DOX detection was from 0.1 to 2.5 μM. Diffusion controlled electrode reaction reveal long l...ife time of the proposed electrode as well high potential for practical application. Developed procedure was successfully applied for the DOX analysis in biological fluids – urine and pharmaceutical formulation. Obtained results clearly indicated that given procedure can be easily implemented for pharmaceutical control and medical analysis, in both, laboratory and field conditions. © 2021 Elsevier B.V.

Кључне речи:
Cytostatic drug / Screen printed diamond electrode / Single drop detection / “Point-of-care” application
Извор:
Diamond and Related Materials, 2021, 113, 108277-
Финансирање / пројекти:
  • EUREKA [E! 13303 MED-BIO-TEST]
  • Grant Agency of the Slovak Republic VEGA [1/0159/20]

DOI: 10.1016/j.diamond.2021.108277

ISSN: 0925-9635

WoS: 000623246800006

Scopus: 2-s2.0-85099612809
[ Google Scholar ]
20
21
URI
https://vinar.vin.bg.ac.rs/handle/123456789/9543
Колекције
  • 070 - Laboratorija za radioizotope
  • Radovi istraživača
Институција/група
Vinča
TY  - JOUR
AU  - Stanković, Dalibor M.
AU  - Milanović, Zorana
AU  - Švorc, Lubomir
AU  - Stanković, Vesna
AU  - Janković, Drina
AU  - Mirković, Marija D.
AU  - Vranješ-Đurić, Sanja
PY  - 2021
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/9543
AB  - This work presents advanced electrochemical platform based on screen printed diamond electrode (SPDE) system for the single drop “point-of-care” testing. Proposed approach was applied for the quantification of doxorubicin hydrochloride (DOX) in biological fluids and pharmaceutical product. Using a single drop (~30 μL) of the tested solution at the electrode surface, DOX showed high electroactivity over a wide range of pHs. In these conditions, single oval shaped, well-defined and pH dependent oxidation peak was observed in the potential range from 0.5 V to 1.3 V. In the reverse scan, two cathodic peaks, were noted – around 0.3 V and – 0.5 V. Similarly, first reduction peak was pH dependent, while second one was independent in the studied range. Experimental conditions for DOX quantification were optimized and natures of the electrode reactions were investigated. Working linear range obtained for DOX detection was from 0.1 to 2.5 μM. Diffusion controlled electrode reaction reveal long life time of the proposed electrode as well high potential for practical application. Developed procedure was successfully applied for the DOX analysis in biological fluids – urine and pharmaceutical formulation. Obtained results clearly indicated that given procedure can be easily implemented for pharmaceutical control and medical analysis, in both, laboratory and field conditions. © 2021 Elsevier B.V.
T2  - Diamond and Related Materials
T1  - Screen printed diamond electrode as efficient “point-of-care” platform for submicromolar determination of cytostatic drug in biological fluids and pharmaceutical product
VL  - 113
SP  - 108277
DO  - 10.1016/j.diamond.2021.108277
ER  - 
@article{
author = "Stanković, Dalibor M. and Milanović, Zorana and Švorc, Lubomir and Stanković, Vesna and Janković, Drina and Mirković, Marija D. and Vranješ-Đurić, Sanja",
year = "2021",
abstract = "This work presents advanced electrochemical platform based on screen printed diamond electrode (SPDE) system for the single drop “point-of-care” testing. Proposed approach was applied for the quantification of doxorubicin hydrochloride (DOX) in biological fluids and pharmaceutical product. Using a single drop (~30 μL) of the tested solution at the electrode surface, DOX showed high electroactivity over a wide range of pHs. In these conditions, single oval shaped, well-defined and pH dependent oxidation peak was observed in the potential range from 0.5 V to 1.3 V. In the reverse scan, two cathodic peaks, were noted – around 0.3 V and – 0.5 V. Similarly, first reduction peak was pH dependent, while second one was independent in the studied range. Experimental conditions for DOX quantification were optimized and natures of the electrode reactions were investigated. Working linear range obtained for DOX detection was from 0.1 to 2.5 μM. Diffusion controlled electrode reaction reveal long life time of the proposed electrode as well high potential for practical application. Developed procedure was successfully applied for the DOX analysis in biological fluids – urine and pharmaceutical formulation. Obtained results clearly indicated that given procedure can be easily implemented for pharmaceutical control and medical analysis, in both, laboratory and field conditions. © 2021 Elsevier B.V.",
journal = "Diamond and Related Materials",
title = "Screen printed diamond electrode as efficient “point-of-care” platform for submicromolar determination of cytostatic drug in biological fluids and pharmaceutical product",
volume = "113",
pages = "108277",
doi = "10.1016/j.diamond.2021.108277"
}
Stanković, D. M., Milanović, Z., Švorc, L., Stanković, V., Janković, D., Mirković, M. D.,& Vranješ-Đurić, S.. (2021). Screen printed diamond electrode as efficient “point-of-care” platform for submicromolar determination of cytostatic drug in biological fluids and pharmaceutical product. in Diamond and Related Materials, 113, 108277.
https://doi.org/10.1016/j.diamond.2021.108277
Stanković DM, Milanović Z, Švorc L, Stanković V, Janković D, Mirković MD, Vranješ-Đurić S. Screen printed diamond electrode as efficient “point-of-care” platform for submicromolar determination of cytostatic drug in biological fluids and pharmaceutical product. in Diamond and Related Materials. 2021;113:108277.
doi:10.1016/j.diamond.2021.108277 .
Stanković, Dalibor M., Milanović, Zorana, Švorc, Lubomir, Stanković, Vesna, Janković, Drina, Mirković, Marija D., Vranješ-Đurić, Sanja, "Screen printed diamond electrode as efficient “point-of-care” platform for submicromolar determination of cytostatic drug in biological fluids and pharmaceutical product" in Diamond and Related Materials, 113 (2021):108277,
https://doi.org/10.1016/j.diamond.2021.108277 . .

DSpace software copyright © 2002-2015  DuraSpace
О репозиторијуму ВинаР | Пошаљите запажања

re3dataOpenAIRERCUB
 

 

Комплетан репозиторијумГрупеАуториНасловиТемеОва институцијаАуториНасловиТеме

Статистика

Преглед статистика

DSpace software copyright © 2002-2015  DuraSpace
О репозиторијуму ВинаР | Пошаљите запажања

re3dataOpenAIRERCUB