Show simple item record

dc.creatorKrneta-Nikolić, Jelena D.
dc.creatorPantelić, Gordana K.
dc.creatorŽivanović, Miloš Z.
dc.creatorRajačić, Milica
dc.creatorTodorović, Dragana
dc.date.accessioned2018-03-03T14:56:24Z
dc.date.available2018-03-03T14:56:24Z
dc.date.issued2014
dc.identifier.issn0144-8420 (print)
dc.identifier.issn1742-3406 (electronic)
dc.identifier.urihttp://vinar.vin.bg.ac.rs/handle/123456789/7047
dc.description.abstractThe charcoal canister method of radon measurement according to US Environment Protection Agency protocol 520/5-87-005 is widely used for screening. This method is based on radon adsorption on coal and measurement of gamma radiation of radon daughters. For the purpose of gamma spectrometry, appropriate efficiency calibration of the measuring system must be performed. The most usual method of calibration is using standard canister, a sealed canister with the same matrix and geometry as the canisters used for measurements, but with the known activity of radon. In the absence of standard canister, a different method of efficiency calibration has to be implemented. This study presents the results of efficiency calibration using the EFFTRAN efficiency transfer software. Efficiency was calculated using a soil matrix cylindrical secondary reference material as a starting point. Calculated efficiency is then compared with the one obtained using standard canister and applied to a realistic measurement in order to evaluate the results of the efficiency transfer.en
dc.relationinfo:eu-repo/grantAgreement/MESTD/Integrated and Interdisciplinary Research (IIR or III)/43009/RS//
dc.rightsrestrictedAccessen
dc.sourceRadiation Protection Dosimetryen
dc.titleComparison of Two Methods for High Purity Germanium Detector Efficiency Calibration for Charcoal Canister Radon Measurementen
dc.typearticleen
dcterms.abstractРајачић Милица; Пантелиц, Г.; Крнета-Николић Јелена; Живановић Милош; Тодоровић Драгана;
dc.citation.volume162
dc.citation.issue1-2
dc.citation.spage47
dc.citation.epage51
dc.identifier.wos000347733500012
dc.identifier.doi10.1093/rpd/ncu216
dc.citation.rankM23
dc.identifier.pmid25377751
dc.description.other2nd International Conference on Radiation and Dosimetry in Various Fields of Research RAD and The 2nd East European Radon Symposium (SEERAS), May 27-30, 2014, Nis, Serbiaen
dc.identifier.scopus2-s2.0-84922420970


Files in this item

FilesSizeFormatView

There are no files associated with this item.

This item appears in the following Collection(s)