Приказ основних података о документу

dc.creatorVassileva, J.
dc.creatorRehani, M.
dc.creatorKostova-Lefterova, D.
dc.creatorAl-Naemi, H. M.
dc.creatorAl Suwaidi, J. S.
dc.creatorAranđić, Danijela
dc.creatorBashier, E. H. O.
dc.creatorKodlulovich Renha, S.
dc.creatorEl-Nachef, L.
dc.creatorAguilar, J. G.
dc.creatorGershan, Vesna
dc.creatorGershkevitsh, E.
dc.creatorGruppetta, E.
dc.creatorHustuc, A.
dc.creatorJauhari, A.
dc.creatorKharita, Mohammad Hassan
dc.creatorKhelassi-Toutaoui, N.
dc.creatorKhosravi, H. R.
dc.creatorKhoury, H.
dc.creatorKralik, I.
dc.creatorMahere, S.
dc.creatorMazuoliene, J.
dc.creatorMora, P.
dc.creatorMuhogora, W.
dc.creatorMuthuvelu, P.
dc.creatorNikodemova, D.
dc.creatorNovak, L.
dc.creatorPallewatte, A.
dc.creatorPekarovic, D.
dc.creatorShaaban, M.
dc.creatorShelly, E.
dc.creatorStepanyan, K.
dc.creatorThelsy, N.
dc.creatorVisrutaratna, P.
dc.creatorZaman, A.
dc.date.accessioned2018-03-01T16:18:39Z
dc.date.available2018-03-01T16:18:39Z
dc.date.issued2015
dc.identifier.issn0144-8420
dc.identifier.issn1742-3406
dc.identifier.urihttps://vinar.vin.bg.ac.rs/handle/123456789/668
dc.description.abstractThe article reports from the largest international dose survey in paediatric computed tomography (CT) in 32 countries and proposes international diagnostic reference levels (DRLs) in terms of computed tomography dose index (CTDIvol) and dose length product (DLP). It also assesses whether mean or median values of individual facilities should be used. A total of 6115 individual patient data were recorded among four age groups: LT 1 y, GT 1-5 y, GT 5-10 y and GT 10-15 y. CTDIw, CTDIvol and DLP from the CT console were recorded in dedicated forms together with patient data and technical parameters. Statistical analysis was performed, and international DRLs were established at rounded 75th percentile values of distribution of median values from all CT facilities. The study presents evidence in favour of using median rather than mean of patient dose indices as the representative of typical local dose in a facility, and for establishing DRLs as third quartile of median values. International DRLs were established for paediatric CT examinations for routine head, chest and abdomen in the four age groups. DRLs for CTDIvol are similar to the reference values from other published reports, with some differences for chest and abdomen CT. Higher variations were observed between DLP values, based on a survey of whole multi-phase exams. It may be noted that other studies in literature were based on single phase only. DRLs reported in this article can be used in countries without sufficient medical physics support to identify non-optimised practice. Recommendations to improve the accuracy and importance of future surveys are provided.en
dc.rightsrestrictedAccessen
dc.sourceRadiation Protection Dosimetryen
dc.titleA Study to Establish International Diagnostic Reference Levels for Paediatric Computed Tomographyen
dc.typearticleen
dc.rights.licenseARR
dcterms.abstractГерсхкевитсх, Е.; Ел-Нацхеф, Л.; Висрутаратна, П.; Басхиер, Е. Х. О.; Aл Суwаиди, Ј. С.; Кхосрави, Х. Р.; Пекаровиц, Д.; Хустуц, A.; Кхарита, Мохаммад Хассан; Никодемова, Д.; Мора, П.; Заман, A.; Кралик, И.; Вассилева, Ј.; Рехани, М.; Aл-Наеми, Х. М.; Aгуилар, Ј. Г.; Јаухари, A.; Кхоурy, Х.; Мухогора, W.; Новак, Л.; Степанyан, К.; Тхелсy, Н.; Кодлуловицх Ренха, С.; Герсхан, В.; Aрандјиц, Д.; Костова-Лефтерова, Д.; Махере, С.; Схеллy, Е.; Кхеласси-Тоутаоуи, Н.; Группетта, Е.; Схаабан, М.; Мазуолиене, Ј.; Паллеwатте, A.; Мутхувелу, П.;
dc.citation.volume165
dc.citation.issue1-4
dc.citation.spage70
dc.citation.epage80
dc.identifier.wos000358449300019
dc.identifier.doi10.1093/rpd/ncv116
dc.citation.rankM23
dc.identifier.pmid25836685