Uticaj kalibracije efikasnosti na rezultate merenja poluprovodnčkim HPGe spektrometrom

The influence of efficiency calibration on measument results with a semiconductor HPGe spectrometer

Thumbnail
2025
Преузимање 🢃
Conference article [PDF] (1.128Mb)
Аутори
Đurašević, Mirjana
Kandić, Aleksandar
Samolov, Aleksandra
Milanović, Tamara
Čeliković, Igor
Конференцијски прилог (Објављена верзија)
,
© Institut za nuklearne nauke „Vinča“ - Institut od nacionalnog značaja za Republiku Srbiju, Univerzitet u Beogradu
Метаподаци
Приказ свих података о документу
Апстракт
Kvantitativno određivanje sadržaja radionuklida gama spektrometrijom zahteva poznavanje efikasnosti detekcije gama zračenja, pa se otuda velika pažnja posvećuje kalibraciji efikasnosti HPGe spektrometra. U ovom radu prikazani su rezultati eksperimenta u kome je korišćen uzorak filter papira poznate aktivnosti. Uzorak je meren na poluprovodničkom HPGe spektrometru u dve različite geometrije (direktno na detektoru i u geometriji plastičnog nosača). Za određivanje aktivnost 134Cs korišćene su dve krive efikasnosti: kriva efikasnosti standarda filter papira koji je postavljen u plastičnom nosaču i kriva efikasnosti dobijena transferom efikasnosti uz pomoć softverskog paketa EFFTRAN. Dobijeni rezultati su pokazali da je najveće odstupanje (14,3 %) dobijeno u slučaju kada se razlikuju geometrije merenog uzorka i standarda. Druge dve analizirane situacije obuhvataju korišćenje transfera efikasnosti kako bi se prevazišla razlika u geometriji, i slučaj u kome su iste geometrije uzorka i standar...da. Dobijeni rezultati aktivnosti 134Cs, u obe situacije, pokazali su prihvatljivo odstupanje od referentne vrednosti (3,5 % i 4,4 %).

Quantitative determination of radionuclide content by gamma spectrometry requires knowledge of the efficiency of gamma radiation detection, so great attention is paid to the calibration of the efficiency of the HPGe spectrometer. This paper presents the results of an experiment in which a filter paper sample of known activity was used. The sample was measured on a semiconductor HPGe spectrometer in two different geometries (directly on the detector and in the geometry of a plastic support). 134Cs activities were obtained using two efficiency curves: the efficiency curve of the filter paper standard placed in a plastic holder and the efficiency curves obtained by efficiency transfer using the EFFTRAN software. The obtained results showed that the largest deviation (14,3 %) was obtained in the case when the geometries of the measured sample and the standard differ. The other two analyzed situations include the use of efficiency transfer to overcome the difference in geometry, and the cas...e when the geometries of the sample and the standard are the same. The obtained results of 134Cs activity, in both situations, showed an acceptable deviation from the reference value (3,5 % and 4,4 %).

Извор:
XXXIII симпозијум ДЗЗСЦГ : зборник радова, 2025, 258-263
Издавач:
  • Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе
Финансирање / пројекти:
  • Министарство науке, технолошког развоја и иновација Републике Србије, институционално финансирање - 200017 (Универзитет у Београду, Институт за нуклеарне науке Винча, Београд-Винча) (RS-MESTD-inst-2020-200017)
Напомена:
  • XXXIII симпозијум ДЗЗСЦГ Друштва за заштиту од зрачења Србије и Црне Горе : 1-3. октобар 2025. године.

ISBN: 978-86-7306-178-8

[ Google Scholar ]
Handle
https://hdl.handle.net/21.15107/rcub_vinar_15955
URI
https://vinar.vin.bg.ac.rs/handle/123456789/15955
Колекције
  • Zbornici Društva za zaštitu od zračenja Srbije i Crne Gore
  • Radovi istraživača
Институција/група
Vinča
TY  - CONF
AU  - Đurašević, Mirjana
AU  - Kandić, Aleksandar
AU  - Samolov, Aleksandra
AU  - Milanović, Tamara
AU  - Čeliković, Igor
PY  - 2025
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/15955
AB  - Kvantitativno određivanje sadržaja radionuklida gama spektrometrijom zahteva poznavanje efikasnosti detekcije gama zračenja, pa se otuda velika pažnja posvećuje kalibraciji efikasnosti HPGe spektrometra. U ovom radu prikazani su rezultati eksperimenta u kome je korišćen uzorak filter papira poznate aktivnosti. Uzorak je meren na poluprovodničkom HPGe spektrometru u dve različite geometrije (direktno na detektoru i u geometriji plastičnog nosača). Za određivanje aktivnost 134Cs korišćene su dve krive efikasnosti: kriva efikasnosti standarda filter papira koji je postavljen u plastičnom nosaču i kriva efikasnosti dobijena transferom efikasnosti uz pomoć softverskog paketa EFFTRAN. Dobijeni rezultati su pokazali da je najveće odstupanje (14,3 %) dobijeno u slučaju kada se razlikuju geometrije merenog uzorka i standarda. Druge dve analizirane situacije obuhvataju korišćenje transfera efikasnosti kako bi se prevazišla razlika u geometriji, i slučaj u kome su iste geometrije uzorka i standarda. Dobijeni rezultati aktivnosti 134Cs, u obe situacije, pokazali su prihvatljivo odstupanje od referentne vrednosti (3,5 % i 4,4 %).
AB  - Quantitative determination of radionuclide content by gamma spectrometry requires knowledge of the efficiency of gamma radiation detection, so great attention is paid to the calibration of the efficiency of the HPGe spectrometer. This paper presents the results of an experiment in which a filter paper sample of known activity was used. The sample was measured on a semiconductor HPGe spectrometer in two different geometries (directly on the detector and in the geometry of a plastic support). 134Cs activities were obtained using two efficiency curves: the efficiency curve of the filter paper standard placed in a plastic holder and the efficiency curves obtained by efficiency transfer using the EFFTRAN software. The obtained results showed that the largest deviation (14,3 %) was obtained in the case when the geometries of the measured sample and the standard differ. The other two analyzed situations include the use of efficiency transfer to overcome the difference in geometry, and the case when the geometries of the sample and the standard are the same. The obtained results of 134Cs activity, in both situations, showed an acceptable deviation from the reference value (3,5 % and 4,4 %).
PB  - Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе
C3  - XXXIII симпозијум ДЗЗСЦГ : зборник радова
T1  - Uticaj kalibracije efikasnosti na rezultate merenja poluprovodnčkim HPGe spektrometrom
T1  - The influence of efficiency calibration on measument results with a semiconductor HPGe spectrometer
SP  - 258
EP  - 263
UR  - https://hdl.handle.net/21.15107/rcub_vinar_15955
ER  - 
@conference{
author = "Đurašević, Mirjana and Kandić, Aleksandar and Samolov, Aleksandra and Milanović, Tamara and Čeliković, Igor",
year = "2025",
abstract = "Kvantitativno određivanje sadržaja radionuklida gama spektrometrijom zahteva poznavanje efikasnosti detekcije gama zračenja, pa se otuda velika pažnja posvećuje kalibraciji efikasnosti HPGe spektrometra. U ovom radu prikazani su rezultati eksperimenta u kome je korišćen uzorak filter papira poznate aktivnosti. Uzorak je meren na poluprovodničkom HPGe spektrometru u dve različite geometrije (direktno na detektoru i u geometriji plastičnog nosača). Za određivanje aktivnost 134Cs korišćene su dve krive efikasnosti: kriva efikasnosti standarda filter papira koji je postavljen u plastičnom nosaču i kriva efikasnosti dobijena transferom efikasnosti uz pomoć softverskog paketa EFFTRAN. Dobijeni rezultati su pokazali da je najveće odstupanje (14,3 %) dobijeno u slučaju kada se razlikuju geometrije merenog uzorka i standarda. Druge dve analizirane situacije obuhvataju korišćenje transfera efikasnosti kako bi se prevazišla razlika u geometriji, i slučaj u kome su iste geometrije uzorka i standarda. Dobijeni rezultati aktivnosti 134Cs, u obe situacije, pokazali su prihvatljivo odstupanje od referentne vrednosti (3,5 % i 4,4 %)., Quantitative determination of radionuclide content by gamma spectrometry requires knowledge of the efficiency of gamma radiation detection, so great attention is paid to the calibration of the efficiency of the HPGe spectrometer. This paper presents the results of an experiment in which a filter paper sample of known activity was used. The sample was measured on a semiconductor HPGe spectrometer in two different geometries (directly on the detector and in the geometry of a plastic support). 134Cs activities were obtained using two efficiency curves: the efficiency curve of the filter paper standard placed in a plastic holder and the efficiency curves obtained by efficiency transfer using the EFFTRAN software. The obtained results showed that the largest deviation (14,3 %) was obtained in the case when the geometries of the measured sample and the standard differ. The other two analyzed situations include the use of efficiency transfer to overcome the difference in geometry, and the case when the geometries of the sample and the standard are the same. The obtained results of 134Cs activity, in both situations, showed an acceptable deviation from the reference value (3,5 % and 4,4 %).",
publisher = "Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе",
journal = "XXXIII симпозијум ДЗЗСЦГ : зборник радова",
title = "Uticaj kalibracije efikasnosti na rezultate merenja poluprovodnčkim HPGe spektrometrom, The influence of efficiency calibration on measument results with a semiconductor HPGe spectrometer",
pages = "258-263",
url = "https://hdl.handle.net/21.15107/rcub_vinar_15955"
}
Đurašević, M., Kandić, A., Samolov, A., Milanović, T.,& Čeliković, I.. (2025). Uticaj kalibracije efikasnosti na rezultate merenja poluprovodnčkim HPGe spektrometrom. in XXXIII симпозијум ДЗЗСЦГ : зборник радова
Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе., 258-263.
https://hdl.handle.net/21.15107/rcub_vinar_15955
Đurašević M, Kandić A, Samolov A, Milanović T, Čeliković I. Uticaj kalibracije efikasnosti na rezultate merenja poluprovodnčkim HPGe spektrometrom. in XXXIII симпозијум ДЗЗСЦГ : зборник радова. 2025;:258-263.
https://hdl.handle.net/21.15107/rcub_vinar_15955 .
Đurašević, Mirjana, Kandić, Aleksandar, Samolov, Aleksandra, Milanović, Tamara, Čeliković, Igor, "Uticaj kalibracije efikasnosti na rezultate merenja poluprovodnčkim HPGe spektrometrom" in XXXIII симпозијум ДЗЗСЦГ : зборник радова (2025):258-263,
https://hdl.handle.net/21.15107/rcub_vinar_15955 .

DSpace software copyright © 2002-2015  DuraSpace
О репозиторијуму ВинаР | Пошаљите запажања

re3dataOpenAIRERCUB
 

 

Комплетан репозиторијумГрупеАуториНасловиТемеОва институцијаАуториНасловиТеме

Статистика

Преглед статистика

DSpace software copyright © 2002-2015  DuraSpace
О репозиторијуму ВинаР | Пошаљите запажања

re3dataOpenAIRERCUB