TLD-100 post-irradiation fading characteristics according to IEC 62387:2020 standard

Thumbnail
2021
Преузимање 🢃
Conference article [PDF] (2.663Mb)
Аутори
Stanković-Petrović, Jelena S.
Krajinović, Marko
Kržanović, Nikola
Živanović, Miloš
Kojić, Andrea
Božović, Predrag
Конференцијски прилог (Објављена верзија)
Метаподаци
Приказ свих података о документу
Апстракт
The results of the post-irradiation fading of whole body dosemeters based on two TLD-100TM (Thermo Scientific™ Harshaw™, USA) detectors are presented. The dosemeters are regularly used by accredited individual monitoring service (IMS) at Vinca Institute of Nuclear Science (VINS), Belgrade, Serbia. The testing of post-irradiation fading was carried out according to International Electrotechnical Commission (IEC) 62387:2020 standard. The irradiations are done in S-Cs137 field at secondary standard dosimetry laboratory (SSDL) at VINS, according to International Organization for Standardization document ISO 4037-3:2019. The chosen reference personal dose equivalent value was 3 mSv. The research was carried out in the period from October 2020 to February 2021, and it lasted 128 days. The frequency of irradiations was approximately 7 days, while 3 periods between irradiations were longer than 10 days, due to COVID-19 pandemic. The irradiations were performed at different dates; thus, all dos...e readouts were done on the same day to prevent influence of reader’s instabilities. The irradiated dosemeters were stored at the same room where the average temperature was nearly 20°C. There were 14 groups with 6 dosemeters for irradiation and 2 dosemeters for natural background radiation level correction. The reader was Harshaw TLD™ Model 6600 Plus Automated Reader (Thermo Fisher Scientific, USA). The whole glow curve was used for dose estimation (all of 200 channels). The time as influence quantity was considered to be of type F, thus the range of relative response was limiting factor in the analysis. The results showed that the maximum measurement time tmax is 72 days for deep dose (Hp(10)) detector, and 85 days for shallow dose (Hp(0.07)) detector. Thus, the standard’s requirement is satisfied, as it is required minimum of 30 days. The relative response range for all of the 14 groups was from 0.82 to 1.14 and from 0.83 to 1.17, for deep and shallow dose, respectively. The research has limitation as the irradiations were organized aligned to COVID-19 working schedule. One of the consequences of this timetable is lack of 7 days’ time point, thus the values from 16 days’ time point was used as referential.

Извор:
RAD 2021 : 9th International Conference on Radiation in Various Fields of Research : Book of abstracts, 2021, 219-219
Издавач:
  • Niš, Serbia : RAD Centre
Напомена:
  • IX International Conference on Radiation in Various Fields of Research : RAD 2021 : book of abstracts; June 14-18, 2021; Herceg Novi, Montenegro

DOI: 10.21175/rad.abstr.book.2021.31.12

ISBN: 978-86-901150-2-0

[ Google Scholar ]
URI
https://vinar.vin.bg.ac.rs/handle/123456789/11129
Колекције
  • 100 - Laboratorija za zaštitu od zračenja i zaštitu životne sredine
  • Radovi istraživača
Институција/група
Vinča
TY  - CONF
AU  - Stanković-Petrović, Jelena S.
AU  - Krajinović, Marko
AU  - Kržanović, Nikola
AU  - Živanović, Miloš
AU  - Kojić, Andrea
AU  - Božović, Predrag
PY  - 2021
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/11129
AB  - The results of the post-irradiation fading of whole body dosemeters based on two TLD-100TM (Thermo Scientific™ Harshaw™, USA) detectors are presented. The dosemeters are regularly used by accredited individual monitoring service (IMS) at Vinca Institute of Nuclear Science (VINS), Belgrade, Serbia. The testing of post-irradiation fading was carried out according to International Electrotechnical Commission (IEC) 62387:2020 standard. The irradiations are done in S-Cs137 field at secondary standard dosimetry laboratory (SSDL) at VINS, according to International Organization for Standardization document ISO 4037-3:2019. The chosen reference personal dose equivalent value was 3 mSv. The research was carried out in the period from October 2020 to February 2021, and it lasted 128 days. The frequency of irradiations was approximately 7 days, while 3 periods between irradiations were longer than 10 days, due to COVID-19 pandemic. The irradiations were performed at different dates; thus, all dose readouts were done on the same day to prevent influence of reader’s instabilities. The irradiated dosemeters were stored at the same room where the average temperature was nearly 20°C. There were 14 groups with 6 dosemeters for irradiation and 2 dosemeters for natural background radiation level correction. The reader was Harshaw TLD™ Model 6600 Plus Automated Reader (Thermo Fisher Scientific, USA). The whole glow curve was used for dose estimation (all of 200 channels). The time as influence quantity was considered to be of type F, thus the range of relative response was limiting factor in the analysis. The results showed that the maximum measurement time tmax is 72 days for deep dose (Hp(10)) detector, and 85 days for shallow dose (Hp(0.07)) detector. Thus, the standard’s requirement is satisfied, as it is required minimum of 30 days. The relative response range for all of the 14 groups was from 0.82 to 1.14 and from 0.83 to 1.17, for deep and shallow dose, respectively. The research has limitation as the irradiations were organized aligned to COVID-19 working schedule. One of the consequences of this timetable is lack of 7 days’ time point, thus the values from 16 days’ time point was used as referential.
PB  - Niš, Serbia : RAD Centre
C3  - RAD 2021 : 9th International Conference on Radiation in Various Fields of Research : Book of abstracts
T1  - TLD-100 post-irradiation fading characteristics according to IEC 62387:2020 standard
SP  - 219
EP  - 219
DO  - 10.21175/rad.abstr.book.2021.31.12
ER  - 
@conference{
author = "Stanković-Petrović, Jelena S. and Krajinović, Marko and Kržanović, Nikola and Živanović, Miloš and Kojić, Andrea and Božović, Predrag",
year = "2021",
abstract = "The results of the post-irradiation fading of whole body dosemeters based on two TLD-100TM (Thermo Scientific™ Harshaw™, USA) detectors are presented. The dosemeters are regularly used by accredited individual monitoring service (IMS) at Vinca Institute of Nuclear Science (VINS), Belgrade, Serbia. The testing of post-irradiation fading was carried out according to International Electrotechnical Commission (IEC) 62387:2020 standard. The irradiations are done in S-Cs137 field at secondary standard dosimetry laboratory (SSDL) at VINS, according to International Organization for Standardization document ISO 4037-3:2019. The chosen reference personal dose equivalent value was 3 mSv. The research was carried out in the period from October 2020 to February 2021, and it lasted 128 days. The frequency of irradiations was approximately 7 days, while 3 periods between irradiations were longer than 10 days, due to COVID-19 pandemic. The irradiations were performed at different dates; thus, all dose readouts were done on the same day to prevent influence of reader’s instabilities. The irradiated dosemeters were stored at the same room where the average temperature was nearly 20°C. There were 14 groups with 6 dosemeters for irradiation and 2 dosemeters for natural background radiation level correction. The reader was Harshaw TLD™ Model 6600 Plus Automated Reader (Thermo Fisher Scientific, USA). The whole glow curve was used for dose estimation (all of 200 channels). The time as influence quantity was considered to be of type F, thus the range of relative response was limiting factor in the analysis. The results showed that the maximum measurement time tmax is 72 days for deep dose (Hp(10)) detector, and 85 days for shallow dose (Hp(0.07)) detector. Thus, the standard’s requirement is satisfied, as it is required minimum of 30 days. The relative response range for all of the 14 groups was from 0.82 to 1.14 and from 0.83 to 1.17, for deep and shallow dose, respectively. The research has limitation as the irradiations were organized aligned to COVID-19 working schedule. One of the consequences of this timetable is lack of 7 days’ time point, thus the values from 16 days’ time point was used as referential.",
publisher = "Niš, Serbia : RAD Centre",
journal = "RAD 2021 : 9th International Conference on Radiation in Various Fields of Research : Book of abstracts",
title = "TLD-100 post-irradiation fading characteristics according to IEC 62387:2020 standard",
pages = "219-219",
doi = "10.21175/rad.abstr.book.2021.31.12"
}
Stanković-Petrović, J. S., Krajinović, M., Kržanović, N., Živanović, M., Kojić, A.,& Božović, P.. (2021). TLD-100 post-irradiation fading characteristics according to IEC 62387:2020 standard. in RAD 2021 : 9th International Conference on Radiation in Various Fields of Research : Book of abstracts
Niš, Serbia : RAD Centre., 219-219.
https://doi.org/10.21175/rad.abstr.book.2021.31.12
Stanković-Petrović JS, Krajinović M, Kržanović N, Živanović M, Kojić A, Božović P. TLD-100 post-irradiation fading characteristics according to IEC 62387:2020 standard. in RAD 2021 : 9th International Conference on Radiation in Various Fields of Research : Book of abstracts. 2021;:219-219.
doi:10.21175/rad.abstr.book.2021.31.12 .
Stanković-Petrović, Jelena S., Krajinović, Marko, Kržanović, Nikola, Živanović, Miloš, Kojić, Andrea, Božović, Predrag, "TLD-100 post-irradiation fading characteristics according to IEC 62387:2020 standard" in RAD 2021 : 9th International Conference on Radiation in Various Fields of Research : Book of abstracts (2021):219-219,
https://doi.org/10.21175/rad.abstr.book.2021.31.12 . .

DSpace software copyright © 2002-2015  DuraSpace
О репозиторијуму ВинаР | Пошаљите запажања

re3dataOpenAIRERCUB
 

 

Комплетан репозиторијумГрупеАуториНасловиТемеОва институцијаАуториНасловиТеме

Статистика

Преглед статистика

DSpace software copyright © 2002-2015  DuraSpace
О репозиторијуму ВинаР | Пошаљите запажања

re3dataOpenAIRERCUB