European Radiation Dosimetry Group [EURADOS WG 12 - Dosimetry in Medical Imaging]

Link to this page

European Radiation Dosimetry Group [EURADOS WG 12 - Dosimetry in Medical Imaging]

Authors

Publications

Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures

Jarvinen, Hannu; Farah, Jad; Siiskonen, Teemu; Ciraj-Bjelac, Olivera; Dabin, Jeremie; Carinou, Eleftheria; Domienik-Andrzejewska, Joanna; Kluszczynski, Dariusz; Knežević, Željka; Kopec, Renata; Majer, Marija; Malchair, Francoise; Negri, Anna; Pankowski, Piotr; Sarmento, Sandra; Trianni, Annalisa

(2018)

TY  - JOUR
AU  - Jarvinen, Hannu
AU  - Farah, Jad
AU  - Siiskonen, Teemu
AU  - Ciraj-Bjelac, Olivera
AU  - Dabin, Jeremie
AU  - Carinou, Eleftheria
AU  - Domienik-Andrzejewska, Joanna
AU  - Kluszczynski, Dariusz
AU  - Knežević, Željka
AU  - Kopec, Renata
AU  - Majer, Marija
AU  - Malchair, Francoise
AU  - Negri, Anna
AU  - Pankowski, Piotr
AU  - Sarmento, Sandra
AU  - Trianni, Annalisa
PY  - 2018
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/7621
AB  - Purpose: The feasibility of setting-up generic, hospital-independent dose alert levels to initiate vigilance on possible skin injuries in interventional procedures was studied for three high-dose procedures (chemoembolization (TACE) of the liver, neuro-embolization (NE) and percutaneous coronary intervention (PCI)) in 9 European countries. Methods: Gafchromic (R) films and thermoluminescent dosimeters (TLD) were used to determine the Maximum Skin Dose (MSD). Correlation of the online dose indicators (fluoroscopy time, kerma- or dose-area product (KAP or DAP) and cumulative air kerma at interventional reference point (K-a,K-r)) with MSD was evaluated and used to establish the alert levels corresponding to a MSD of 2 Gy and 5 Gy. The uncertainties of alert levels in terms of DAP and K-a,K-r, and uncertainty of MSD were calculated. Results: About 20-30% of all MSD values exceeded 2 Gy while only 2-6% exceeded 5 Gy. The correlations suggest that both DAP and K-a,K-r can be used as a dose indicator for alert levels (Pearson correlation coefficient p mostly > 0.8), while fluoroscopy time is not suitable (p mostly < 0.6). Generic alert levels based on DAP (Gy cm(2)) were suggested for MSD of both 2 Gy and 5 Gy (for 5 Gy: TACE 750, PCI 250 and NE 400). The suggested levels are close to the lowest values published in several other studies. The uncertainty of the MSD was estimated to be around 10-15% and of hospital-specific skin dose alert levels about 20-30% (with coverage factor k = 1). Conclusions: The generic alert levels are feasible for some cases but should be used with caution, only as the first approximation, while hospital-specific alert levels are preferred as the final approach.
T2  - Physica Medica
T1  - Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures
VL  - 46
SP  - 67
EP  - 74
DO  - 10.1016/j.ejmp.2018.01.010
ER  - 
@article{
author = "Jarvinen, Hannu and Farah, Jad and Siiskonen, Teemu and Ciraj-Bjelac, Olivera and Dabin, Jeremie and Carinou, Eleftheria and Domienik-Andrzejewska, Joanna and Kluszczynski, Dariusz and Knežević, Željka and Kopec, Renata and Majer, Marija and Malchair, Francoise and Negri, Anna and Pankowski, Piotr and Sarmento, Sandra and Trianni, Annalisa",
year = "2018",
abstract = "Purpose: The feasibility of setting-up generic, hospital-independent dose alert levels to initiate vigilance on possible skin injuries in interventional procedures was studied for three high-dose procedures (chemoembolization (TACE) of the liver, neuro-embolization (NE) and percutaneous coronary intervention (PCI)) in 9 European countries. Methods: Gafchromic (R) films and thermoluminescent dosimeters (TLD) were used to determine the Maximum Skin Dose (MSD). Correlation of the online dose indicators (fluoroscopy time, kerma- or dose-area product (KAP or DAP) and cumulative air kerma at interventional reference point (K-a,K-r)) with MSD was evaluated and used to establish the alert levels corresponding to a MSD of 2 Gy and 5 Gy. The uncertainties of alert levels in terms of DAP and K-a,K-r, and uncertainty of MSD were calculated. Results: About 20-30% of all MSD values exceeded 2 Gy while only 2-6% exceeded 5 Gy. The correlations suggest that both DAP and K-a,K-r can be used as a dose indicator for alert levels (Pearson correlation coefficient p mostly > 0.8), while fluoroscopy time is not suitable (p mostly < 0.6). Generic alert levels based on DAP (Gy cm(2)) were suggested for MSD of both 2 Gy and 5 Gy (for 5 Gy: TACE 750, PCI 250 and NE 400). The suggested levels are close to the lowest values published in several other studies. The uncertainty of the MSD was estimated to be around 10-15% and of hospital-specific skin dose alert levels about 20-30% (with coverage factor k = 1). Conclusions: The generic alert levels are feasible for some cases but should be used with caution, only as the first approximation, while hospital-specific alert levels are preferred as the final approach.",
journal = "Physica Medica",
title = "Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures",
volume = "46",
pages = "67-74",
doi = "10.1016/j.ejmp.2018.01.010"
}
Jarvinen, H., Farah, J., Siiskonen, T., Ciraj-Bjelac, O., Dabin, J., Carinou, E., Domienik-Andrzejewska, J., Kluszczynski, D., Knežević, Ž., Kopec, R., Majer, M., Malchair, F., Negri, A., Pankowski, P., Sarmento, S.,& Trianni, A.. (2018). Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures. in Physica Medica, 46, 67-74.
https://doi.org/10.1016/j.ejmp.2018.01.010
Jarvinen H, Farah J, Siiskonen T, Ciraj-Bjelac O, Dabin J, Carinou E, Domienik-Andrzejewska J, Kluszczynski D, Knežević Ž, Kopec R, Majer M, Malchair F, Negri A, Pankowski P, Sarmento S, Trianni A. Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures. in Physica Medica. 2018;46:67-74.
doi:10.1016/j.ejmp.2018.01.010 .
Jarvinen, Hannu, Farah, Jad, Siiskonen, Teemu, Ciraj-Bjelac, Olivera, Dabin, Jeremie, Carinou, Eleftheria, Domienik-Andrzejewska, Joanna, Kluszczynski, Dariusz, Knežević, Željka, Kopec, Renata, Majer, Marija, Malchair, Francoise, Negri, Anna, Pankowski, Piotr, Sarmento, Sandra, Trianni, Annalisa, "Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures" in Physica Medica, 46 (2018):67-74,
https://doi.org/10.1016/j.ejmp.2018.01.010 . .
16
14
16