IAEA Coordinated Research Project E24024, entitled ‘Evaluation of the Dosimetry Needs and Practices for the Update of the Code of Practice for Dosimetry in Diagnostic Radiology (TRS-457)

Link to this page

IAEA Coordinated Research Project E24024, entitled ‘Evaluation of the Dosimetry Needs and Practices for the Update of the Code of Practice for Dosimetry in Diagnostic Radiology (TRS-457)

Authors

Publications

Evaluation of diagnostic radiology detector performance in reference mammography radiation fields

Kržanović, Nikola; Kojić, Andrea; Bakrač, Luka; Živanović, Miloš

(Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе, 2023)

TY  - CONF
AU  - Kržanović, Nikola
AU  - Kojić, Andrea
AU  - Bakrač, Luka
AU  - Živanović, Miloš
PY  - 2023
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/11925
AB  - Recommendations on diagnostic radiology calibrations are defined by the IEC and IAEA documents for molybdenum anode target material and molybdenum primary beam filtration (abbreviated as the RQR-M series). Calibration of diagnostic radiology dosimeters (ionization chambers and X-ray multimeters) can also be performed for other anode/filter combinations, including the tungsten anode / aluminium filtration setup. Performance of dosimeters can significantly vary in radiation fields defined by different anode materials and primary beam filtrations, where different spectra and first half-value layers are obtained for the same nominal X-ray tube voltages. Therefore it is important to calibrate or test dosimeters for the range of radiation qualities or conditions in which they will be used. Performance of two such detectors was evaluated.
PB  - Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе
C3  - 32. simpozijum DZZSCG : zbornik radova
T1  - Evaluation of diagnostic radiology detector performance in reference mammography radiation fields
SP  - 369
EP  - 374
UR  - https://hdl.handle.net/21.15107/rcub_vinar_11925
ER  - 
@conference{
author = "Kržanović, Nikola and Kojić, Andrea and Bakrač, Luka and Živanović, Miloš",
year = "2023",
abstract = "Recommendations on diagnostic radiology calibrations are defined by the IEC and IAEA documents for molybdenum anode target material and molybdenum primary beam filtration (abbreviated as the RQR-M series). Calibration of diagnostic radiology dosimeters (ionization chambers and X-ray multimeters) can also be performed for other anode/filter combinations, including the tungsten anode / aluminium filtration setup. Performance of dosimeters can significantly vary in radiation fields defined by different anode materials and primary beam filtrations, where different spectra and first half-value layers are obtained for the same nominal X-ray tube voltages. Therefore it is important to calibrate or test dosimeters for the range of radiation qualities or conditions in which they will be used. Performance of two such detectors was evaluated.",
publisher = "Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе",
journal = "32. simpozijum DZZSCG : zbornik radova",
title = "Evaluation of diagnostic radiology detector performance in reference mammography radiation fields",
pages = "369-374",
url = "https://hdl.handle.net/21.15107/rcub_vinar_11925"
}
Kržanović, N., Kojić, A., Bakrač, L.,& Živanović, M.. (2023). Evaluation of diagnostic radiology detector performance in reference mammography radiation fields. in 32. simpozijum DZZSCG : zbornik radova
Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе., 369-374.
https://hdl.handle.net/21.15107/rcub_vinar_11925
Kržanović N, Kojić A, Bakrač L, Živanović M. Evaluation of diagnostic radiology detector performance in reference mammography radiation fields. in 32. simpozijum DZZSCG : zbornik radova. 2023;:369-374.
https://hdl.handle.net/21.15107/rcub_vinar_11925 .
Kržanović, Nikola, Kojić, Andrea, Bakrač, Luka, Živanović, Miloš, "Evaluation of diagnostic radiology detector performance in reference mammography radiation fields" in 32. simpozijum DZZSCG : zbornik radova (2023):369-374,
https://hdl.handle.net/21.15107/rcub_vinar_11925 .