Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories

Hourdakis, Constantine J.; Bueermann, Ludwig; Ciraj-Bjelac, Olivera; Csete, Istvan; Delis, Harry; Gomola, Igor; Persson, Linda; Novak, Leos; Petkov, Ivailo; Toroi, Paula

(2016)

TY  - JOUR
AU  - Hourdakis, Constantine J.
AU  - Bueermann, Ludwig
AU  - Ciraj-Bjelac, Olivera
AU  - Csete, Istvan
AU  - Delis, Harry
AU  - Gomola, Igor
AU  - Persson, Linda
AU  - Novak, Leos
AU  - Petkov, Ivailo
AU  - Toroi, Paula
PY  - 2016
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/271
AB  - A comparison of calibration results and procedures in terms of air kerma length product, P-KL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the partial irradiation method is the preferred method for the pencil-type IC calibration in terms of P-KL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement ( GT 99%) for both dosimetry quantities (P-KL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed. (C) 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
T2  - Physica Medica
T1  - Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories
VL  - 32
IS  - 1
SP  - 42
EP  - 51
DO  - 10.1016/j.ejmp.2015.09.008
ER  - 
@article{
author = "Hourdakis, Constantine J. and Bueermann, Ludwig and Ciraj-Bjelac, Olivera and Csete, Istvan and Delis, Harry and Gomola, Igor and Persson, Linda and Novak, Leos and Petkov, Ivailo and Toroi, Paula",
year = "2016",
abstract = "A comparison of calibration results and procedures in terms of air kerma length product, P-KL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the partial irradiation method is the preferred method for the pencil-type IC calibration in terms of P-KL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement ( GT 99%) for both dosimetry quantities (P-KL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed. (C) 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.",
journal = "Physica Medica",
title = "Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories",
volume = "32",
number = "1",
pages = "42-51",
doi = "10.1016/j.ejmp.2015.09.008"
}
Hourdakis, C. J., Bueermann, L., Ciraj-Bjelac, O., Csete, I., Delis, H., Gomola, I., Persson, L., Novak, L., Petkov, I.,& Toroi, P.. (2016). Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories. in Physica Medica, 32(1), 42-51.
https://doi.org/10.1016/j.ejmp.2015.09.008
Hourdakis CJ, Bueermann L, Ciraj-Bjelac O, Csete I, Delis H, Gomola I, Persson L, Novak L, Petkov I, Toroi P. Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories. in Physica Medica. 2016;32(1):42-51.
doi:10.1016/j.ejmp.2015.09.008 .
Hourdakis, Constantine J., Bueermann, Ludwig, Ciraj-Bjelac, Olivera, Csete, Istvan, Delis, Harry, Gomola, Igor, Persson, Linda, Novak, Leos, Petkov, Ivailo, Toroi, Paula, "Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories" in Physica Medica, 32, no. 1 (2016):42-51,
https://doi.org/10.1016/j.ejmp.2015.09.008 . .