author = "Kržanović, Nikola and Kojić, Andrea and Bakrač, Luka and Živanović, Miloš",
year = "2023",
abstract = "Recommendations on diagnostic radiology calibrations are defined by the IEC and IAEA documents for molybdenum anode target material and molybdenum primary beam filtration (abbreviated as the RQR-M series). Calibration of diagnostic radiology dosimeters (ionization chambers and X-ray multimeters) can also be performed for other anode/filter combinations, including the tungsten anode / aluminium filtration setup. Performance of dosimeters can significantly vary in radiation fields defined by different anode materials and primary beam filtrations, where different spectra and first half-value layers are obtained for the same nominal X-ray tube voltages. Therefore it is important to calibrate or test dosimeters for the range of radiation qualities or conditions in which they will be used. Performance of two such detectors was evaluated.",
publisher = "Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе",
journal = "32. simpozijum DZZSCG : zbornik radova",
title = "Evaluation of diagnostic radiology detector performance in reference mammography radiation fields",
pages = "369-374",
url = "https://hdl.handle.net/21.15107/rcub_vinar_11925"
}