@article{
author = "Ciraj-Bjelac, Olivera and Beciric, Snezana and Aranđić, Danijela and Košutić, Duško D. and Kovačević, Milojko",
year = "2010",
abstract = "The primary objective of this study is to investigate the actual mammography dose levels in Serbia and establish a baseline dose database for the future screening programme. The mean glandular dose (MGD) was assessed for standard size breast substituted by 45 mm polymethyl methacrylate (PMMA) phantom and imaged under typical clinical conditions at 30 screen-film mammography facilities. Average MGD was (1.9 +/- 1.0) mGy, with a range of 0.12-5.2 mGy. These results were used for the calculation of the initial diagnostic reference levels in mammography in Serbia, which was set to 2.1 mGy for a standard breast. At some facilities, high doses were observed and the assessed MGD was supplemented by a patient dose survey. MGD was measured for 53 women having a diagnostic mammography in cranio-caudal (CC) and mediolateral oblique (MLO) projections at the largest mammography facilities in Serbia. The parameters recorded were: age, compressed breast thickness, tube potential (in kV), tube loading (in mAs) and the MLO projection angle. The average MGD per image was 2.8 mGy for the CC projection and 4.3 mGy for the MW projection. Dose to breast was compared with dose for PMMA phantoms of various sizes, assessed at the same facilities. The results have indicated that phantom dose values can assist in setting optimisation activities in mammography. Both phantom and patient dose values indicated unnecessary high doses in facilities with an extremely high mammography workload. For these facilities, urgent dose-reduction measures and follow-up actions were recommended.",
journal = "Radiation Protection Dosimetry",
title = "Mammography Radiation Dose: Initial Results from Serbia Based on Mean Glandular Dose Assessment for Phantoms and Patients",
volume = "140",
number = "1",
pages = "75-80",
doi = "10.1093/rpd/ncq040"
}