Application and Testing of a New Simple Experimental Set-Up for Thermal Conductivity Measurements of Liquids

Stanimirović, Andrej M.; Živković, Emila M.; Milošević, Nenad D.; Kijevčanin, Mirjana Lj.

(2017)

TY  - JOUR
AU  - Stanimirović, Andrej M.
AU  - Živković, Emila M.
AU  - Milošević, Nenad D.
AU  - Kijevčanin, Mirjana Lj.
PY  - 2017
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/1629
AB  - Transient hot wire method is considered a reliable and precise technique for measuring the thermal conductivity of liquids. The present paper describes a new transient hot wire experimental set-up and its initial testing. The new apparatus was tested by performing thermal conductivity measurements on substances whose reference thermophysical properties data existed in literature, namely on pure toluene and double distilled deionized water. The values of thermal conductivity measured in the temperature range 25 to 45 degrees C deviated +2.2% to +3% from the literature data, while the expanded measurement uncertainty was estimated to be +/- 4%.
T2  - Thermal Science
T1  - Application and Testing of a New Simple Experimental Set-Up for Thermal Conductivity Measurements of Liquids
VL  - 21
IS  - 3
SP  - 1195
EP  - 1202
DO  - 10.2298/TSCI160324219S
ER  - 
@article{
author = "Stanimirović, Andrej M. and Živković, Emila M. and Milošević, Nenad D. and Kijevčanin, Mirjana Lj.",
year = "2017",
abstract = "Transient hot wire method is considered a reliable and precise technique for measuring the thermal conductivity of liquids. The present paper describes a new transient hot wire experimental set-up and its initial testing. The new apparatus was tested by performing thermal conductivity measurements on substances whose reference thermophysical properties data existed in literature, namely on pure toluene and double distilled deionized water. The values of thermal conductivity measured in the temperature range 25 to 45 degrees C deviated +2.2% to +3% from the literature data, while the expanded measurement uncertainty was estimated to be +/- 4%.",
journal = "Thermal Science",
title = "Application and Testing of a New Simple Experimental Set-Up for Thermal Conductivity Measurements of Liquids",
volume = "21",
number = "3",
pages = "1195-1202",
doi = "10.2298/TSCI160324219S"
}
Stanimirović, A. M., Živković, E. M., Milošević, N. D.,& Kijevčanin, M. Lj.. (2017). Application and Testing of a New Simple Experimental Set-Up for Thermal Conductivity Measurements of Liquids. in Thermal Science, 21(3), 1195-1202.
https://doi.org/10.2298/TSCI160324219S
Stanimirović AM, Živković EM, Milošević ND, Kijevčanin ML. Application and Testing of a New Simple Experimental Set-Up for Thermal Conductivity Measurements of Liquids. in Thermal Science. 2017;21(3):1195-1202.
doi:10.2298/TSCI160324219S .
Stanimirović, Andrej M., Živković, Emila M., Milošević, Nenad D., Kijevčanin, Mirjana Lj., "Application and Testing of a New Simple Experimental Set-Up for Thermal Conductivity Measurements of Liquids" in Thermal Science, 21, no. 3 (2017):1195-1202,
https://doi.org/10.2298/TSCI160324219S . .
4
2
4