Type-testing of the TLD system used for the assessing the eye lens doses due to ionising radiation

Stanković Petrović, Jelena; Danilović, Nataša; Kržanović, Nikola; Topalović, Dušan; Živanović, Miloš; Božović, Predrag; Todorović, Nataša; Nikolov, Jovana

(2023)

TY  - JOUR
AU  - Stanković Petrović, Jelena
AU  - Danilović, Nataša
AU  - Kržanović, Nikola
AU  - Topalović, Dušan
AU  - Živanović, Miloš
AU  - Božović, Predrag
AU  - Todorović, Nataša
AU  - Nikolov, Jovana
PY  - 2023
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/11642
AB  - Individual monitoring service at the Department for Radiation and Environmental Protection of Vinca Institute of
Nuclear Sciences performs individual whole-body and extremities monitoring of exposed workers, using thermoluminescent dosimetry (TLD) method. The dosemeters for eye-lens personal dose equivalent measurements,
Hp(3), are not regularly used at the workplaces in Serbia, even though the individual monitoring method is
accredited and available at the dosimetry service. Type testing of Hp(3) dosemeters based on LiF: Mg,Ti, placed in
modified ring holders was done following the changes in ISO 4037:2019 and IEC 62387:2020. The type testing
was done by varying radiation-based influence quantities: photon energies ranging from 33.3 keV up to 1.25
MeV, angle of incidence values of 0◦, 45◦, 60◦, 75◦, dose range from 0.05 mSv to 10 Sv. The results show that the
passive dosimetry monitoring system satisfies IEC 62387:2020 requirements in the mandatory range of influence
quantities, except for the lowest tested energy of 33 keV, when incident radiation angle is ±60◦ and higher.
T2  - Radiation Measurements
T1  - Type-testing of the TLD system used for the assessing the eye lens doses due to ionising radiation
VL  - 168
SP  - 107014
DO  - 10.1016/j.radmeas.2023.107014
ER  - 
@article{
author = "Stanković Petrović, Jelena and Danilović, Nataša and Kržanović, Nikola and Topalović, Dušan and Živanović, Miloš and Božović, Predrag and Todorović, Nataša and Nikolov, Jovana",
year = "2023",
abstract = "Individual monitoring service at the Department for Radiation and Environmental Protection of Vinca Institute of
Nuclear Sciences performs individual whole-body and extremities monitoring of exposed workers, using thermoluminescent dosimetry (TLD) method. The dosemeters for eye-lens personal dose equivalent measurements,
Hp(3), are not regularly used at the workplaces in Serbia, even though the individual monitoring method is
accredited and available at the dosimetry service. Type testing of Hp(3) dosemeters based on LiF: Mg,Ti, placed in
modified ring holders was done following the changes in ISO 4037:2019 and IEC 62387:2020. The type testing
was done by varying radiation-based influence quantities: photon energies ranging from 33.3 keV up to 1.25
MeV, angle of incidence values of 0◦, 45◦, 60◦, 75◦, dose range from 0.05 mSv to 10 Sv. The results show that the
passive dosimetry monitoring system satisfies IEC 62387:2020 requirements in the mandatory range of influence
quantities, except for the lowest tested energy of 33 keV, when incident radiation angle is ±60◦ and higher.",
journal = "Radiation Measurements",
title = "Type-testing of the TLD system used for the assessing the eye lens doses due to ionising radiation",
volume = "168",
pages = "107014",
doi = "10.1016/j.radmeas.2023.107014"
}
Stanković Petrović, J., Danilović, N., Kržanović, N., Topalović, D., Živanović, M., Božović, P., Todorović, N.,& Nikolov, J.. (2023). Type-testing of the TLD system used for the assessing the eye lens doses due to ionising radiation. in Radiation Measurements, 168, 107014.
https://doi.org/10.1016/j.radmeas.2023.107014
Stanković Petrović J, Danilović N, Kržanović N, Topalović D, Živanović M, Božović P, Todorović N, Nikolov J. Type-testing of the TLD system used for the assessing the eye lens doses due to ionising radiation. in Radiation Measurements. 2023;168:107014.
doi:10.1016/j.radmeas.2023.107014 .
Stanković Petrović, Jelena, Danilović, Nataša, Kržanović, Nikola, Topalović, Dušan, Živanović, Miloš, Božović, Predrag, Todorović, Nataša, Nikolov, Jovana, "Type-testing of the TLD system used for the assessing the eye lens doses due to ionising radiation" in Radiation Measurements, 168 (2023):107014,
https://doi.org/10.1016/j.radmeas.2023.107014 . .