Detection limits of the NaI(Tl) shielded HPGe spectrometer

Bikit, Ištvan S.; Slivka, Jaroslav; Krmar, Miodrag; Durcic, Z; Zikic, N; Čonkič, Lj.; Veskovic, M; Aničin, Ivan V.

(1999)

TY  - JOUR
AU  - Bikit, Ištvan S.
AU  - Slivka, Jaroslav
AU  - Krmar, Miodrag
AU  - Durcic, Z
AU  - Zikic, N
AU  - Čonkič, Lj.
AU  - Veskovic, M
AU  - Aničin, Ivan V.
PY  - 1999
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/2220
AB  - The results of a detailed study of the low-level performance of a NaI(Tl) shield added to an iron shielded HPGe spectrometer are presented. Both the slow and the fast anticoincidence gating modes were tested, the slow mode being found better suited for general low-level spectroscopy applications. In long runs the stability of the system in this mode is satisfactory. The anticoincidence action of the NaI(Tl) shield lowers the integral background of the iron shielded HPGe detector in the energy range from 30 keV to 2 MeV by a factor of 6.5, and suppresses the continuum above 150 keV by a factor larger than 10. (C) 1999 Elsevier Science B.V. All rights reserved.
T2  - Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment
T1  - Detection limits of the NaI(Tl) shielded HPGe spectrometer
VL  - 421
IS  - 1-2
SP  - 266
EP  - 271
DO  - 10.1016/S0168-9002(98)01258-3
ER  - 
@article{
author = "Bikit, Ištvan S. and Slivka, Jaroslav and Krmar, Miodrag and Durcic, Z and Zikic, N and Čonkič, Lj. and Veskovic, M and Aničin, Ivan V.",
year = "1999",
abstract = "The results of a detailed study of the low-level performance of a NaI(Tl) shield added to an iron shielded HPGe spectrometer are presented. Both the slow and the fast anticoincidence gating modes were tested, the slow mode being found better suited for general low-level spectroscopy applications. In long runs the stability of the system in this mode is satisfactory. The anticoincidence action of the NaI(Tl) shield lowers the integral background of the iron shielded HPGe detector in the energy range from 30 keV to 2 MeV by a factor of 6.5, and suppresses the continuum above 150 keV by a factor larger than 10. (C) 1999 Elsevier Science B.V. All rights reserved.",
journal = "Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment",
title = "Detection limits of the NaI(Tl) shielded HPGe spectrometer",
volume = "421",
number = "1-2",
pages = "266-271",
doi = "10.1016/S0168-9002(98)01258-3"
}
Bikit, I. S., Slivka, J., Krmar, M., Durcic, Z., Zikic, N., Čonkič, Lj., Veskovic, M.,& Aničin, I. V.. (1999). Detection limits of the NaI(Tl) shielded HPGe spectrometer. in Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment, 421(1-2), 266-271.
https://doi.org/10.1016/S0168-9002(98)01258-3
Bikit IS, Slivka J, Krmar M, Durcic Z, Zikic N, Čonkič L, Veskovic M, Aničin IV. Detection limits of the NaI(Tl) shielded HPGe spectrometer. in Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment. 1999;421(1-2):266-271.
doi:10.1016/S0168-9002(98)01258-3 .
Bikit, Ištvan S., Slivka, Jaroslav, Krmar, Miodrag, Durcic, Z, Zikic, N, Čonkič, Lj., Veskovic, M, Aničin, Ivan V., "Detection limits of the NaI(Tl) shielded HPGe spectrometer" in Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment, 421, no. 1-2 (1999):266-271,
https://doi.org/10.1016/S0168-9002(98)01258-3 . .