Verification of the sampling procedure for radioactivity analysis of waste and industrial material

Rajačić, Milica; Todorović, Dragana; Andrić, Velibor; Nikolić, Jelena K.; Janković, Marija M.; Pantelić, Gordana K.; Vukanac, Ivana

(2023)

TY  - JOUR
AU  - Rajačić, Milica
AU  - Todorović, Dragana
AU  - Andrić, Velibor
AU  - Nikolić, Jelena K.
AU  - Janković, Marija M.
AU  - Pantelić, Gordana K.
AU  - Vukanac, Ivana
PY  - 2023
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/11065
AB  - In accordance with ISO/IEC 17025:2017, the testing laboratory needs to verify its ability to properly perform methods. For radiological testing, the sampling does not affect the results directly, but it has to ensure that sample adequately represents the tested material. To verify the procedure, a sampling of red mud and bauxite ore was conducted. All samples were measured in identical geometry by HPGe spectrometer. The counting rates per unit mass in the recorded spectra were compared. For each measurement series, the mean and standard deviations of respected peaks were calculated, as well as average and standard deviation of all series. The obtained results of each individual series were considered satisfactory, i.e. sampling procedure ensures the representativeness of the bulk material, if it is within ± two standard deviations of the average of the mean values.
T2  - Radiation Protection Dosimetry
T1  - Verification of the sampling procedure for radioactivity analysis of waste and industrial material
VL  - 199
IS  - 8-9
SP  - 941
EP  - 946
DO  - 10.1093/rpd/ncad095
ER  - 
@article{
author = "Rajačić, Milica and Todorović, Dragana and Andrić, Velibor and Nikolić, Jelena K. and Janković, Marija M. and Pantelić, Gordana K. and Vukanac, Ivana",
year = "2023",
abstract = "In accordance with ISO/IEC 17025:2017, the testing laboratory needs to verify its ability to properly perform methods. For radiological testing, the sampling does not affect the results directly, but it has to ensure that sample adequately represents the tested material. To verify the procedure, a sampling of red mud and bauxite ore was conducted. All samples were measured in identical geometry by HPGe spectrometer. The counting rates per unit mass in the recorded spectra were compared. For each measurement series, the mean and standard deviations of respected peaks were calculated, as well as average and standard deviation of all series. The obtained results of each individual series were considered satisfactory, i.e. sampling procedure ensures the representativeness of the bulk material, if it is within ± two standard deviations of the average of the mean values.",
journal = "Radiation Protection Dosimetry",
title = "Verification of the sampling procedure for radioactivity analysis of waste and industrial material",
volume = "199",
number = "8-9",
pages = "941-946",
doi = "10.1093/rpd/ncad095"
}
Rajačić, M., Todorović, D., Andrić, V., Nikolić, J. K., Janković, M. M., Pantelić, G. K.,& Vukanac, I.. (2023). Verification of the sampling procedure for radioactivity analysis of waste and industrial material. in Radiation Protection Dosimetry, 199(8-9), 941-946.
https://doi.org/10.1093/rpd/ncad095
Rajačić M, Todorović D, Andrić V, Nikolić JK, Janković MM, Pantelić GK, Vukanac I. Verification of the sampling procedure for radioactivity analysis of waste and industrial material. in Radiation Protection Dosimetry. 2023;199(8-9):941-946.
doi:10.1093/rpd/ncad095 .
Rajačić, Milica, Todorović, Dragana, Andrić, Velibor, Nikolić, Jelena K., Janković, Marija M., Pantelić, Gordana K., Vukanac, Ivana, "Verification of the sampling procedure for radioactivity analysis of waste and industrial material" in Radiation Protection Dosimetry, 199, no. 8-9 (2023):941-946,
https://doi.org/10.1093/rpd/ncad095 . .