@article{
author = "Travar, M. and Nikolov, Jovana and Todorović, N. and Vraničar, A. and Völgyesi, P. and Kirchknopf, P. and Čeliković, Igor T. and Milanović, Tamara J. and Joković, D.",
year = "2023",
abstract = "Presented study describes the optimization method of an HPGe detector through implementation of Geant4 toolkit. The optimized model was verified through comparison with experimentally obtained data using a set of point-like radioactive calibration sources. Acquired results displayed good agreement with the experimental data that falls under an average relative deviation of the order of ~ 2% within the energy range of 53–1836 keV. Additionally, in order to test the validity of the code it was also applied to a different detection equipment where an average relative deviation of the order of ~ 1.8% was achieved within the energy range of 121–1112 keV. © 2023, Akadémiai Kiadó, Budapest, Hungary.",
journal = "Journal of Radioanalytical and Nuclear Chemistry",
title = "Detailed optimization procedure of an HPGe detector using Geant4 toolkit",
volume = "332",
number = "3",
pages = "817-828",
doi = "10.1007/s10967-023-08810-x"
}