Estimation of absorbed doses to target and healthy tissues during cone-beam ct radiotherapy imaging of the chest

Marjanović, Milana S.; Tóth, Árpad; Petrović, Borislava; Gencel, Ivan V.

(2021)

TY  - JOUR
AU  - Marjanović, Milana S.
AU  - Tóth, Árpad
AU  - Petrović, Borislava
AU  - Gencel, Ivan V.
PY  - 2021
UR  - https://vinar.vin.bg.ac.rs/handle/123456789/10218
AB  - The aim of this study was to assess volumetric imaging dose and image quality during radiotherapy in case of Hodgkin lymphoma. The absorbed doses were measured using ionization chamber in thorax phantom. For the same setup, Monte Carlo simulations were performed. The image quality was assessed using Catphan phantom. The absorbed doses ranged from 4.4–7.7 mGy (full rotation) and 1.1–3 mGy (partial rotation). Doses from simulation ranged from 6.8–15.6 mGy (full rotation) and 1.9–10.4 mGy (partial rotation). The image quality tests have shown differences between clinical protocols. The imaging doses were smaller than prescribed doses, but shouldn’t be neglected. © 2021, Publishing House of the Romanian Academy. All rights reserved.
T2  - Romanian Journal of Physics
T1  - Estimation of absorbed doses to target and healthy tissues during cone-beam ct radiotherapy imaging of the chest
VL  - 66
IS  - 9-10
UR  - https://hdl.handle.net/21.15107/rcub_vinar_10218
ER  - 
@article{
author = "Marjanović, Milana S. and Tóth, Árpad and Petrović, Borislava and Gencel, Ivan V.",
year = "2021",
abstract = "The aim of this study was to assess volumetric imaging dose and image quality during radiotherapy in case of Hodgkin lymphoma. The absorbed doses were measured using ionization chamber in thorax phantom. For the same setup, Monte Carlo simulations were performed. The image quality was assessed using Catphan phantom. The absorbed doses ranged from 4.4–7.7 mGy (full rotation) and 1.1–3 mGy (partial rotation). Doses from simulation ranged from 6.8–15.6 mGy (full rotation) and 1.9–10.4 mGy (partial rotation). The image quality tests have shown differences between clinical protocols. The imaging doses were smaller than prescribed doses, but shouldn’t be neglected. © 2021, Publishing House of the Romanian Academy. All rights reserved.",
journal = "Romanian Journal of Physics",
title = "Estimation of absorbed doses to target and healthy tissues during cone-beam ct radiotherapy imaging of the chest",
volume = "66",
number = "9-10",
url = "https://hdl.handle.net/21.15107/rcub_vinar_10218"
}
Marjanović, M. S., Tóth, Á., Petrović, B.,& Gencel, I. V.. (2021). Estimation of absorbed doses to target and healthy tissues during cone-beam ct radiotherapy imaging of the chest. in Romanian Journal of Physics, 66(9-10).
https://hdl.handle.net/21.15107/rcub_vinar_10218
Marjanović MS, Tóth Á, Petrović B, Gencel IV. Estimation of absorbed doses to target and healthy tissues during cone-beam ct radiotherapy imaging of the chest. in Romanian Journal of Physics. 2021;66(9-10).
https://hdl.handle.net/21.15107/rcub_vinar_10218 .
Marjanović, Milana S., Tóth, Árpad, Petrović, Borislava, Gencel, Ivan V., "Estimation of absorbed doses to target and healthy tissues during cone-beam ct radiotherapy imaging of the chest" in Romanian Journal of Physics, 66, no. 9-10 (2021),
https://hdl.handle.net/21.15107/rcub_vinar_10218 .