Experimental determination of the HPGe spectrometer efficiency curve
Nema prikaza
Autori
Vukanac, IvanaĐurašević, Mirjana M.
Kandić, Aleksandar
Novković, Dušan N.
Nađđerđ, Laslo
Milošević, Zoran
Članak u časopisu
Metapodaci
Prikaz svih podataka o dokumentuApstrakt
A review of the calibration procedures of the semiconductor HPGe spectrometer is presented in this paper. Calibration standards were prepared using the standardized radioactive solution of the common monoenergetic radionuclides mixture and standardized multigamma Eu-152 solution. The matrix materials were spiked using the activated carbon, and homogenized mechanically or by suspension in C2H5OH. Experimentally determined efficiency curves were compared with those obtained by Monte Carlo simulation. (C) 2008 Elsevier Ltd. All rights reserved.
Izvor:
Applied Radiation and Isotopes, 2008, 66, 6-7, 792-795Napomena:
- 16th International Conference on Radionuclide Metrology and Its Applications (ICRM 2007), Sep 03-07, 2007, Cape Town, South Africa
DOI: 10.1016/j.apradiso.2008.02.039
ISSN: 0969-8043
PubMed: 18343144
WoS: 000256203000024
Scopus: 2-s2.0-43049181894
Kolekcije
Institucija/grupa
VinčaTY - JOUR AU - Vukanac, Ivana AU - Đurašević, Mirjana M. AU - Kandić, Aleksandar AU - Novković, Dušan N. AU - Nađđerđ, Laslo AU - Milošević, Zoran PY - 2008 UR - https://vinar.vin.bg.ac.rs/handle/123456789/6756 AB - A review of the calibration procedures of the semiconductor HPGe spectrometer is presented in this paper. Calibration standards were prepared using the standardized radioactive solution of the common monoenergetic radionuclides mixture and standardized multigamma Eu-152 solution. The matrix materials were spiked using the activated carbon, and homogenized mechanically or by suspension in C2H5OH. Experimentally determined efficiency curves were compared with those obtained by Monte Carlo simulation. (C) 2008 Elsevier Ltd. All rights reserved. T2 - Applied Radiation and Isotopes T1 - Experimental determination of the HPGe spectrometer efficiency curve VL - 66 IS - 6-7 SP - 792 EP - 795 DO - 10.1016/j.apradiso.2008.02.039 ER -
@article{ author = "Vukanac, Ivana and Đurašević, Mirjana M. and Kandić, Aleksandar and Novković, Dušan N. and Nađđerđ, Laslo and Milošević, Zoran", year = "2008", abstract = "A review of the calibration procedures of the semiconductor HPGe spectrometer is presented in this paper. Calibration standards were prepared using the standardized radioactive solution of the common monoenergetic radionuclides mixture and standardized multigamma Eu-152 solution. The matrix materials were spiked using the activated carbon, and homogenized mechanically or by suspension in C2H5OH. Experimentally determined efficiency curves were compared with those obtained by Monte Carlo simulation. (C) 2008 Elsevier Ltd. All rights reserved.", journal = "Applied Radiation and Isotopes", title = "Experimental determination of the HPGe spectrometer efficiency curve", volume = "66", number = "6-7", pages = "792-795", doi = "10.1016/j.apradiso.2008.02.039" }
Vukanac, I., Đurašević, M. M., Kandić, A., Novković, D. N., Nađđerđ, L.,& Milošević, Z.. (2008). Experimental determination of the HPGe spectrometer efficiency curve. in Applied Radiation and Isotopes, 66(6-7), 792-795. https://doi.org/10.1016/j.apradiso.2008.02.039
Vukanac I, Đurašević MM, Kandić A, Novković DN, Nađđerđ L, Milošević Z. Experimental determination of the HPGe spectrometer efficiency curve. in Applied Radiation and Isotopes. 2008;66(6-7):792-795. doi:10.1016/j.apradiso.2008.02.039 .
Vukanac, Ivana, Đurašević, Mirjana M., Kandić, Aleksandar, Novković, Dušan N., Nađđerđ, Laslo, Milošević, Zoran, "Experimental determination of the HPGe spectrometer efficiency curve" in Applied Radiation and Isotopes, 66, no. 6-7 (2008):792-795, https://doi.org/10.1016/j.apradiso.2008.02.039 . .