Приказ основних података о документу

dc.creatorVesić, D.
dc.creatorPuzović, Jovan M.
dc.creatorAničin, Ivan V.
dc.date.accessioned2018-03-01T18:41:55Z
dc.date.available2018-03-01T18:41:55Z
dc.date.issued1999
dc.identifier.issn0969-8043
dc.identifier.urihttps://vinar.vin.bg.ac.rs/handle/123456789/2242
dc.description.abstractWe show that for cylindrical voluminous sources in gamma-ray spectroscopy the relative error of the spectral intensity of a low-energy spectral line, as the height of the source increases, saturates faster than the spectral intensity itself. This significantly reduces the quantity of the required sample material as compared to that based on the usual considerations of the spectral intensity only. We also demonstrate that there exists a weakly pronounced height of the source (the sourer size) which minimizes the relative error of the intensity of a given low-energy spectral line, Due to the smallness of the effect the benefit of introducing the concept of the truly optimum size of the source is more on a psychological side, one should feel easier using an optimum quantity of the source material than merely a sufficient one, as is otherwise the case. (C) 1999 Elsevier Science Ltd. All rights reserved.en
dc.rightsrestrictedAccessen
dc.rightsrestrictedAccess
dc.sourceApplied Radiation and Isotopesen
dc.subjectGamma-spectroscopy
dc.subjectLarge sources
dc.subjectErrors
dc.subjectOptimum quantity
dc.titleTowards the definition of the optimum height of voluminous cylindrical sources in gamma-ray spectroscopyen
dc.typearticleen
dc.rights.licenseARR
dcterms.abstractВесиц, Д; Пузовиц, Ј; Aничин Иван В;
dc.citation.volume50
dc.citation.issue6
dc.citation.spage999
dc.citation.epage1001
dc.identifier.wos000079835300004
dc.identifier.doi10.1016/S0969-8043(98)00172-9
dc.citation.rankM21
dc.type.versionpublishedVersion
dc.identifier.scopus2-s2.0-0032899571