Приказ основних података о документу
Intercomparison and performance assessment of radionuclide calibrators used in nuclear medicine departments in Serbia
dc.creator | Živanović, Miloš Z. | |
dc.creator | Vukanac, Ivana | |
dc.creator | Samac, Jelena | |
dc.creator | Antić, Vojislav | |
dc.creator | Jeremić, Marija | |
dc.creator | Janković, Drina | |
dc.creator | Kandić, Aleksandar | |
dc.creator | Vargas, Clarita Saldarriaga | |
dc.creator | Struelens, Lara | |
dc.creator | Ciraj-Bjelac, Olivera | |
dc.date.accessioned | 2021-12-16T07:58:29Z | |
dc.date.available | 2021-12-16T07:58:29Z | |
dc.date.issued | 2022 | |
dc.identifier.issn | 0969-8043 | |
dc.identifier.uri | https://vinar.vin.bg.ac.rs/handle/123456789/10016 | |
dc.description.abstract | The purpose of this work is to assess accuracy and compare the performance of radionuclide calibrators (RNCs) used in nuclear medicine departments in Serbia. Testing of the RNCs included verification of measurement accuracy, as well as analysis of routinely used quality control protocols, by using the certified radioactivity standards (57Co, 137Cs). RNCs performances were assessed with 99mTc through comparison of reference value for radionuclide activity and RNC measurements. Results of the intercomparison revealed that RNCs, 15 in total, are accurate within 10% in vial geometry and within 15% in syringe geometry. Most of them showed similar performance. The results revealed that container geometry is an important influencing parameter in the accuracy of activity measurement. Obtained results indicate a need for regular calibration and implementation of Quality Control program in order to achieve and maintain the accuracy of activity measurements in nuclear medicine. | en |
dc.language | en | |
dc.relation | Ministry of Education, Science & Technological Development of Serbia | |
dc.relation | Belgian Federal Ministry | |
dc.rights | restrictedAccess | |
dc.source | Applied Radiation and Isotopes | |
dc.subject | Accuracy | en |
dc.subject | Intercomparison study | en |
dc.subject | Linearity | en |
dc.subject | Performance testing | en |
dc.subject | Precision | en |
dc.subject | Radionuclide calibrators | en |
dc.subject | Stability | en |
dc.title | Intercomparison and performance assessment of radionuclide calibrators used in nuclear medicine departments in Serbia | en |
dc.type | article | en |
dc.rights.license | ARR | |
dc.citation.volume | 179 | |
dc.citation.spage | 110013 | |
dc.identifier.wos | 000718045400007 | |
dc.identifier.doi | 10.1016/j.apradiso.2021.110013 | |
dc.identifier.pmid | 34741955 | |
dc.type.version | publishedVersion | |
dc.identifier.scopus | 2-s2.0-85118560727 |
Документи
Овај документ се појављује у следећим колекцијама
-
011 - Laboratorija za nuklearnu i plazma fiziku
Department of Nuclear and Plasma Physics -
Radovi istraživača
Researchers' publications