Notes on the calibration of routine dosimeters in radiation processing
Конференцијски прилог (Објављена верзија)
,
Society of Physical Chemists of Serbia
Метаподаци
Приказ свих података о документуАпстракт
The essential prerequisite of radiation dosimetry is to provide quality assurance and
documentation that the irradiation procedure has been carried out according to
specifications requirement of the correct calibration of the chosen dosimetry
system. In the Radiation Plant at the Vinča Institute, we compared two
recommended protocols in calibration of dosimetry systems in radiation
processing: (i) by using standardized routine dosimeters (ethanol-chlorobenzene
ECB) from the reference laboratory and (ii) by in-plant calibration with the alanine
transfer dosimeters. The analysis of results showed that our in-plant calibration is
as good as when standardized dosimeters are used, thus validating our irradiation
geometry in the irradiation process.
Извор:
Physical chemistry 2010 : 10th international conference on fundamental and applied aspects of physical chemistry, 2010Издавач:
- Society of Physical Chemists of Serbia
Колекције
Институција/група
VinčaTY - CONF AU - Šećerov, Bojana AU - Bačić, Goran PY - 2010 UR - https://vinar.vin.bg.ac.rs/handle/123456789/9322 AB - The essential prerequisite of radiation dosimetry is to provide quality assurance and documentation that the irradiation procedure has been carried out according to specifications requirement of the correct calibration of the chosen dosimetry system. In the Radiation Plant at the Vinča Institute, we compared two recommended protocols in calibration of dosimetry systems in radiation processing: (i) by using standardized routine dosimeters (ethanol-chlorobenzene ECB) from the reference laboratory and (ii) by in-plant calibration with the alanine transfer dosimeters. The analysis of results showed that our in-plant calibration is as good as when standardized dosimeters are used, thus validating our irradiation geometry in the irradiation process. PB - Society of Physical Chemists of Serbia C3 - Physical chemistry 2010 : 10th international conference on fundamental and applied aspects of physical chemistry T1 - Notes on the calibration of routine dosimeters in radiation processing UR - https://hdl.handle.net/21.15107/rcub_vinar_9322 ER -
@conference{ author = "Šećerov, Bojana and Bačić, Goran", year = "2010", abstract = "The essential prerequisite of radiation dosimetry is to provide quality assurance and documentation that the irradiation procedure has been carried out according to specifications requirement of the correct calibration of the chosen dosimetry system. In the Radiation Plant at the Vinča Institute, we compared two recommended protocols in calibration of dosimetry systems in radiation processing: (i) by using standardized routine dosimeters (ethanol-chlorobenzene ECB) from the reference laboratory and (ii) by in-plant calibration with the alanine transfer dosimeters. The analysis of results showed that our in-plant calibration is as good as when standardized dosimeters are used, thus validating our irradiation geometry in the irradiation process.", publisher = "Society of Physical Chemists of Serbia", journal = "Physical chemistry 2010 : 10th international conference on fundamental and applied aspects of physical chemistry", title = "Notes on the calibration of routine dosimeters in radiation processing", url = "https://hdl.handle.net/21.15107/rcub_vinar_9322" }
Šećerov, B.,& Bačić, G.. (2010). Notes on the calibration of routine dosimeters in radiation processing. in Physical chemistry 2010 : 10th international conference on fundamental and applied aspects of physical chemistry Society of Physical Chemists of Serbia.. https://hdl.handle.net/21.15107/rcub_vinar_9322
Šećerov B, Bačić G. Notes on the calibration of routine dosimeters in radiation processing. in Physical chemistry 2010 : 10th international conference on fundamental and applied aspects of physical chemistry. 2010;. https://hdl.handle.net/21.15107/rcub_vinar_9322 .
Šećerov, Bojana, Bačić, Goran, "Notes on the calibration of routine dosimeters in radiation processing" in Physical chemistry 2010 : 10th international conference on fundamental and applied aspects of physical chemistry (2010), https://hdl.handle.net/21.15107/rcub_vinar_9322 .