Characterization of dosimeters used in radiation processing
Конференцијски прилог (Објављена верзија)
,
Society of Physical Chemists of Serbia
Метаподаци
Приказ свих података о документуАпстракт
Aérial L-alanine pellet dosimeter is characterized as the dispersion of dose
measurement values in order to determine the level of uncertainty caused by
dosimeter itself. The results of repeated measurements indicate that the uncertainty
of 1.2% could be attributed to the homogeneity of alanine in dosimeter pellets. It
was obtained by analyzing the results of measurements of the different dosimeters
from the same batch that the uncertainty due to difference in dosimeter mass is 1%.
The total uncertainty due to dosimeter itself is 1.5%.
Извор:
Physical chemistry 2012 : 11th international conference on fundamental and applied aspects of physical chemistry, 2012, 1, 412-414Издавач:
- Society of Physical Chemists of Serbia
Напомена:
- Physical chemistry 2012 : 11th international conference on fundamental and applied aspects of physical chemistry; Belgrade (Serbia); 24-28 September 2012
Колекције
Институција/група
VinčaTY - CONF AU - Šećerov, Bojana AU - Janković, Ivana A. AU - Dramićanin, Miroslav PY - 2012 UR - https://vinar.vin.bg.ac.rs/handle/123456789/9307 AB - Aérial L-alanine pellet dosimeter is characterized as the dispersion of dose measurement values in order to determine the level of uncertainty caused by dosimeter itself. The results of repeated measurements indicate that the uncertainty of 1.2% could be attributed to the homogeneity of alanine in dosimeter pellets. It was obtained by analyzing the results of measurements of the different dosimeters from the same batch that the uncertainty due to difference in dosimeter mass is 1%. The total uncertainty due to dosimeter itself is 1.5%. PB - Society of Physical Chemists of Serbia C3 - Physical chemistry 2012 : 11th international conference on fundamental and applied aspects of physical chemistry T1 - Characterization of dosimeters used in radiation processing VL - 1 SP - 412 EP - 414 UR - https://hdl.handle.net/21.15107/rcub_vinar_9307 ER -
@conference{ author = "Šećerov, Bojana and Janković, Ivana A. and Dramićanin, Miroslav", year = "2012", abstract = "Aérial L-alanine pellet dosimeter is characterized as the dispersion of dose measurement values in order to determine the level of uncertainty caused by dosimeter itself. The results of repeated measurements indicate that the uncertainty of 1.2% could be attributed to the homogeneity of alanine in dosimeter pellets. It was obtained by analyzing the results of measurements of the different dosimeters from the same batch that the uncertainty due to difference in dosimeter mass is 1%. The total uncertainty due to dosimeter itself is 1.5%.", publisher = "Society of Physical Chemists of Serbia", journal = "Physical chemistry 2012 : 11th international conference on fundamental and applied aspects of physical chemistry", title = "Characterization of dosimeters used in radiation processing", volume = "1", pages = "412-414", url = "https://hdl.handle.net/21.15107/rcub_vinar_9307" }
Šećerov, B., Janković, I. A.,& Dramićanin, M.. (2012). Characterization of dosimeters used in radiation processing. in Physical chemistry 2012 : 11th international conference on fundamental and applied aspects of physical chemistry Society of Physical Chemists of Serbia., 1, 412-414. https://hdl.handle.net/21.15107/rcub_vinar_9307
Šećerov B, Janković IA, Dramićanin M. Characterization of dosimeters used in radiation processing. in Physical chemistry 2012 : 11th international conference on fundamental and applied aspects of physical chemistry. 2012;1:412-414. https://hdl.handle.net/21.15107/rcub_vinar_9307 .
Šećerov, Bojana, Janković, Ivana A., Dramićanin, Miroslav, "Characterization of dosimeters used in radiation processing" in Physical chemistry 2012 : 11th international conference on fundamental and applied aspects of physical chemistry, 1 (2012):412-414, https://hdl.handle.net/21.15107/rcub_vinar_9307 .