Upgrading of the CAPRICE type ECR ion source
Authors
Efremov, AndreyBogomolov, Sergey
Bekhterev, Vladimir
Dobrosavljević, Aleksandar S.
Vujović, Velibor V.
Nešković, Nebojša B.
Trajić, Ivan
Conference object (Published version)
,
© 2018 Author(s)
Metadata
Show full item recordAbstract
The CAPRICE-type ECR ion source mVINIS has been upgraded by increasing its magnetic field to improve a plasma confinement and thereby enhance the source performance. This modification made it also possible to increase the internal diameter of the plasma chamber and to replace the coaxial microwave input by a waveguide. Some major subsystems such as: the vacuum system, the microwave system, the gas inlet system, the solid substance inlet system, and the control system have been also refurbished. All these improvements have resulted in a substantial increase of ion beam currents, especially in the case of high charge states, with the operation of the ion source proven to be stable and reproducible. This modification can be applied to other CAPRICE-type ion sources. © 2018 Author(s).
Keywords:
vacuum systems / plasma confinement / ion sourcesSource:
AIP Conference Proceedings, 2018, 2011, 040017-Note:
- 17th International Conference on Ion Sources 2018; Geneva's International Conference Centre Geneva; Switzerland; 15 September 2017 through 20 September 2017; Code 139926
DOI: 10.1063/1.5053291
ISBN: 9780735417274
ISSN: 0094-243X
WoS: 000457042700049
Scopus: 2-s2.0-85054170771
URI
http://aip.scitation.org/doi/abs/10.1063/1.5053291https://vinar.vin.bg.ac.rs/handle/123456789/7900
Collections
Institution/Community
VinčaTY - CONF AU - Efremov, Andrey AU - Bogomolov, Sergey AU - Bekhterev, Vladimir AU - Dobrosavljević, Aleksandar S. AU - Vujović, Velibor V. AU - Nešković, Nebojša B. AU - Trajić, Ivan PY - 2018 UR - http://aip.scitation.org/doi/abs/10.1063/1.5053291 UR - https://vinar.vin.bg.ac.rs/handle/123456789/7900 AB - The CAPRICE-type ECR ion source mVINIS has been upgraded by increasing its magnetic field to improve a plasma confinement and thereby enhance the source performance. This modification made it also possible to increase the internal diameter of the plasma chamber and to replace the coaxial microwave input by a waveguide. Some major subsystems such as: the vacuum system, the microwave system, the gas inlet system, the solid substance inlet system, and the control system have been also refurbished. All these improvements have resulted in a substantial increase of ion beam currents, especially in the case of high charge states, with the operation of the ion source proven to be stable and reproducible. This modification can be applied to other CAPRICE-type ion sources. © 2018 Author(s). C3 - AIP Conference Proceedings T1 - Upgrading of the CAPRICE type ECR ion source VL - 2011 SP - 040017 DO - 10.1063/1.5053291 ER -
@conference{ author = "Efremov, Andrey and Bogomolov, Sergey and Bekhterev, Vladimir and Dobrosavljević, Aleksandar S. and Vujović, Velibor V. and Nešković, Nebojša B. and Trajić, Ivan", year = "2018", abstract = "The CAPRICE-type ECR ion source mVINIS has been upgraded by increasing its magnetic field to improve a plasma confinement and thereby enhance the source performance. This modification made it also possible to increase the internal diameter of the plasma chamber and to replace the coaxial microwave input by a waveguide. Some major subsystems such as: the vacuum system, the microwave system, the gas inlet system, the solid substance inlet system, and the control system have been also refurbished. All these improvements have resulted in a substantial increase of ion beam currents, especially in the case of high charge states, with the operation of the ion source proven to be stable and reproducible. This modification can be applied to other CAPRICE-type ion sources. © 2018 Author(s).", journal = "AIP Conference Proceedings", title = "Upgrading of the CAPRICE type ECR ion source", volume = "2011", pages = "040017", doi = "10.1063/1.5053291" }
Efremov, A., Bogomolov, S., Bekhterev, V., Dobrosavljević, A. S., Vujović, V. V., Nešković, N. B.,& Trajić, I.. (2018). Upgrading of the CAPRICE type ECR ion source. in AIP Conference Proceedings, 2011, 040017. https://doi.org/10.1063/1.5053291
Efremov A, Bogomolov S, Bekhterev V, Dobrosavljević AS, Vujović VV, Nešković NB, Trajić I. Upgrading of the CAPRICE type ECR ion source. in AIP Conference Proceedings. 2018;2011:040017. doi:10.1063/1.5053291 .
Efremov, Andrey, Bogomolov, Sergey, Bekhterev, Vladimir, Dobrosavljević, Aleksandar S., Vujović, Velibor V., Nešković, Nebojša B., Trajić, Ivan, "Upgrading of the CAPRICE type ECR ion source" in AIP Conference Proceedings, 2011 (2018):040017, https://doi.org/10.1063/1.5053291 . .