Application of PHOTON simulation software on calibration of HPGe detectors
Апстракт
One of the major difficulties in gamma spectrometry of voluminous environmental samples is the efficiency calibration of the detectors used for the measurement. The direct measurement of different calibration sources, containing isolated gamma-ray emitters within the energy range of interest, and subsequent fitting to a parametric function, is the most accurate and at the same time most complicated and time consuming method of efficiency calibration. Many other methods are developed in time, some of them using Monte Carlo simulation. One of such methods is a dedicated and user-friendly program PHOTON, developed to simulate the passage of photons through different media with different geometries. This program was used for efficiency calibration of three HPGe detectors, readily used in Laboratory for Environment and Radiation Protection of the Institute for Nuclear Sciences Vinca, Belgrade, Serbia. The simulation produced the spectral response of the detectors for fixed energy and for di...fferent sample geometries and matrices. Thus obtained efficiencies were compared to the values obtained by the measurement of the secondary reference materials and to the results obtained by GEANT4 simulation, in order to establish whether the simulated values agree with the experimental ones. To further analyze the results, a realistic measurement of the materials provided by the IAEA within different interlaboratory proficiency tests, was performed. The activities obtained using simulated efficiencies were compared to the reference values provided by the organizer. A good agreement in the mid energy section of the spectrum was obtained, while for low energies the lack of some parameters in the simulation libraries proved to produce unacceptable discrepancies. (C) 2015 Elsevier B.V. All rights reserved.
Кључне речи:
HPGe / Calibration / PHOTONИзвор:
Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment, 2015, 799, 159-165Финансирање / пројекти:
- Нове технологије за мониторинг и заштиту животног окружења од штетних хемијских супстанци и радијационог оптерећења (RS-MESTD-Integrated and Interdisciplinary Research (IIR or III)-43009)
DOI: 10.1016/j.nima.2015.08.002
ISSN: 0168-9002; 1872-9576
WoS: 000361877300024
Scopus: 2-s2.0-84939798445
Колекције
Институција/група
VinčaTY - JOUR AU - Krneta-Nikolić, Jelena D. AU - Puzović, Jovan M. AU - Todorović, Dragana AU - Rajačić, Milica PY - 2015 UR - https://vinar.vin.bg.ac.rs/handle/123456789/753 AB - One of the major difficulties in gamma spectrometry of voluminous environmental samples is the efficiency calibration of the detectors used for the measurement. The direct measurement of different calibration sources, containing isolated gamma-ray emitters within the energy range of interest, and subsequent fitting to a parametric function, is the most accurate and at the same time most complicated and time consuming method of efficiency calibration. Many other methods are developed in time, some of them using Monte Carlo simulation. One of such methods is a dedicated and user-friendly program PHOTON, developed to simulate the passage of photons through different media with different geometries. This program was used for efficiency calibration of three HPGe detectors, readily used in Laboratory for Environment and Radiation Protection of the Institute for Nuclear Sciences Vinca, Belgrade, Serbia. The simulation produced the spectral response of the detectors for fixed energy and for different sample geometries and matrices. Thus obtained efficiencies were compared to the values obtained by the measurement of the secondary reference materials and to the results obtained by GEANT4 simulation, in order to establish whether the simulated values agree with the experimental ones. To further analyze the results, a realistic measurement of the materials provided by the IAEA within different interlaboratory proficiency tests, was performed. The activities obtained using simulated efficiencies were compared to the reference values provided by the organizer. A good agreement in the mid energy section of the spectrum was obtained, while for low energies the lack of some parameters in the simulation libraries proved to produce unacceptable discrepancies. (C) 2015 Elsevier B.V. All rights reserved. T2 - Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment T1 - Application of PHOTON simulation software on calibration of HPGe detectors VL - 799 SP - 159 EP - 165 DO - 10.1016/j.nima.2015.08.002 ER -
@article{ author = "Krneta-Nikolić, Jelena D. and Puzović, Jovan M. and Todorović, Dragana and Rajačić, Milica", year = "2015", abstract = "One of the major difficulties in gamma spectrometry of voluminous environmental samples is the efficiency calibration of the detectors used for the measurement. The direct measurement of different calibration sources, containing isolated gamma-ray emitters within the energy range of interest, and subsequent fitting to a parametric function, is the most accurate and at the same time most complicated and time consuming method of efficiency calibration. Many other methods are developed in time, some of them using Monte Carlo simulation. One of such methods is a dedicated and user-friendly program PHOTON, developed to simulate the passage of photons through different media with different geometries. This program was used for efficiency calibration of three HPGe detectors, readily used in Laboratory for Environment and Radiation Protection of the Institute for Nuclear Sciences Vinca, Belgrade, Serbia. The simulation produced the spectral response of the detectors for fixed energy and for different sample geometries and matrices. Thus obtained efficiencies were compared to the values obtained by the measurement of the secondary reference materials and to the results obtained by GEANT4 simulation, in order to establish whether the simulated values agree with the experimental ones. To further analyze the results, a realistic measurement of the materials provided by the IAEA within different interlaboratory proficiency tests, was performed. The activities obtained using simulated efficiencies were compared to the reference values provided by the organizer. A good agreement in the mid energy section of the spectrum was obtained, while for low energies the lack of some parameters in the simulation libraries proved to produce unacceptable discrepancies. (C) 2015 Elsevier B.V. All rights reserved.", journal = "Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment", title = "Application of PHOTON simulation software on calibration of HPGe detectors", volume = "799", pages = "159-165", doi = "10.1016/j.nima.2015.08.002" }
Krneta-Nikolić, J. D., Puzović, J. M., Todorović, D.,& Rajačić, M.. (2015). Application of PHOTON simulation software on calibration of HPGe detectors. in Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment, 799, 159-165. https://doi.org/10.1016/j.nima.2015.08.002
Krneta-Nikolić JD, Puzović JM, Todorović D, Rajačić M. Application of PHOTON simulation software on calibration of HPGe detectors. in Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment. 2015;799:159-165. doi:10.1016/j.nima.2015.08.002 .
Krneta-Nikolić, Jelena D., Puzović, Jovan M., Todorović, Dragana, Rajačić, Milica, "Application of PHOTON simulation software on calibration of HPGe detectors" in Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment, 799 (2015):159-165, https://doi.org/10.1016/j.nima.2015.08.002 . .