Non-invasive respiratory monitoring using long-period fiber grating sensors
Аутори
Petrovic, M. D.Petrović, Jovana S.
Daničić, Aleksandar
Vukcevic, M.
Bojović, Boško
Hadžievski, Ljupčo
Allsop, Thomas P.
Lloyd, G.
Webb, David J.
Чланак у часопису
Метаподаци
Приказ свих података о документуАпстракт
In non-invasive ventilation, continuous monitoring of respiratory volumes is essential. Here, we present a method for the measurement of respiratory volumes by a single fiber-grating sensor of bending and provide the proof-of-principle by applying a calibration-test measurement procedure on a set of 18 healthy volunteers. Results establish a linear correlation between a change in lung volume and the corresponding change in a local thorax curvature. They also show good sensor accuracy in measurements of tidal and minute respiratory volumes for different types of breathing. The proposed technique does not rely on the air flow through an oronasal mask or the observation of chest movement by a clinician, which distinguishes it from the current clinical practice. (C) 2014 Optical Society of America
Извор:
Biomedical Optics Express, 2014, 5, 4, 1136-1144Финансирање / пројекти:
- Фотоника микро и нано структурних материјала (RS-MESTD-Integrated and Interdisciplinary Research (IIR or III)-45010)
DOI: 10.1364/BOE.5.001136
ISSN: 2156-7085
PubMed: 24761295
WoS: 000334078100013
Scopus: 2-s2.0-84896974052
Колекције
Институција/група
VinčaTY - JOUR AU - Petrovic, M. D. AU - Petrović, Jovana S. AU - Daničić, Aleksandar AU - Vukcevic, M. AU - Bojović, Boško AU - Hadžievski, Ljupčo AU - Allsop, Thomas P. AU - Lloyd, G. AU - Webb, David J. PY - 2014 UR - https://vinar.vin.bg.ac.rs/handle/123456789/5946 AB - In non-invasive ventilation, continuous monitoring of respiratory volumes is essential. Here, we present a method for the measurement of respiratory volumes by a single fiber-grating sensor of bending and provide the proof-of-principle by applying a calibration-test measurement procedure on a set of 18 healthy volunteers. Results establish a linear correlation between a change in lung volume and the corresponding change in a local thorax curvature. They also show good sensor accuracy in measurements of tidal and minute respiratory volumes for different types of breathing. The proposed technique does not rely on the air flow through an oronasal mask or the observation of chest movement by a clinician, which distinguishes it from the current clinical practice. (C) 2014 Optical Society of America T2 - Biomedical Optics Express T1 - Non-invasive respiratory monitoring using long-period fiber grating sensors VL - 5 IS - 4 SP - 1136 EP - 1144 DO - 10.1364/BOE.5.001136 ER -
@article{ author = "Petrovic, M. D. and Petrović, Jovana S. and Daničić, Aleksandar and Vukcevic, M. and Bojović, Boško and Hadžievski, Ljupčo and Allsop, Thomas P. and Lloyd, G. and Webb, David J.", year = "2014", abstract = "In non-invasive ventilation, continuous monitoring of respiratory volumes is essential. Here, we present a method for the measurement of respiratory volumes by a single fiber-grating sensor of bending and provide the proof-of-principle by applying a calibration-test measurement procedure on a set of 18 healthy volunteers. Results establish a linear correlation between a change in lung volume and the corresponding change in a local thorax curvature. They also show good sensor accuracy in measurements of tidal and minute respiratory volumes for different types of breathing. The proposed technique does not rely on the air flow through an oronasal mask or the observation of chest movement by a clinician, which distinguishes it from the current clinical practice. (C) 2014 Optical Society of America", journal = "Biomedical Optics Express", title = "Non-invasive respiratory monitoring using long-period fiber grating sensors", volume = "5", number = "4", pages = "1136-1144", doi = "10.1364/BOE.5.001136" }
Petrovic, M. D., Petrović, J. S., Daničić, A., Vukcevic, M., Bojović, B., Hadžievski, L., Allsop, T. P., Lloyd, G.,& Webb, D. J.. (2014). Non-invasive respiratory monitoring using long-period fiber grating sensors. in Biomedical Optics Express, 5(4), 1136-1144. https://doi.org/10.1364/BOE.5.001136
Petrovic MD, Petrović JS, Daničić A, Vukcevic M, Bojović B, Hadžievski L, Allsop TP, Lloyd G, Webb DJ. Non-invasive respiratory monitoring using long-period fiber grating sensors. in Biomedical Optics Express. 2014;5(4):1136-1144. doi:10.1364/BOE.5.001136 .
Petrovic, M. D., Petrović, Jovana S., Daničić, Aleksandar, Vukcevic, M., Bojović, Boško, Hadžievski, Ljupčo, Allsop, Thomas P., Lloyd, G., Webb, David J., "Non-invasive respiratory monitoring using long-period fiber grating sensors" in Biomedical Optics Express, 5, no. 4 (2014):1136-1144, https://doi.org/10.1364/BOE.5.001136 . .