Hydroxyapatite Coatings on Cp-Titanium Grade-2 Surfaces Prepared with Plasma Spraying
Nema prikaza
Autori
Rudolf, RebekaStamenkovic, Dragoslav
Aleksić, Zoran
Jenko, Monika
Dordevic, Igor
Todorović, Aleksandar
Jokanović, Vukoman R.
Raic, Karlo T.
Članak u časopisu (Objavljena verzija)
Metapodaci
Prikaz svih podataka o dokumentuApstrakt
Thin hydroxyapatite coatings were produced on Cp-Titanium Grade-2 samples, with new high-voltage pulse-power equipment PJ-100 (Plasma Jet, Serbia) in order to get a more stable implant structure appropriate for further clinical applications. A comparative analysis of differently prepared surfaces of the Cp-Titanium Grade-2 samples was done before the hydroxyapatite was applied. Microstructural observation of the modified hydroxyapatite/implant surface was done using scanning-electron-microscopy imaging and Auger electron spectroscopy, with the aim of detecting the morphology and the elements contained in the new surfaces of the samples. The results confirmed that the surface of Cp-Titanium Grade-2 modified with hydroxyapatite is very similar to the bone structure.
Ključne reči:
Cp-Ti2 material / hydroxyapatite (HA) / plasma-spray coating / characterizationIzvor:
Materials and Technology = Materiali in Tehnologije, 2015, 49, 1, 81-86Finansiranje / projekti:
- Hemijsko i strukturno dizajniranje nanomaterijala za primenu u medicini i inženjerstvu tkiva (RS-MESTD-Basic Research (BR or ON)-172026)
- EUREKA Programme [CELL-TI E!5831]
Kolekcije
Institucija/grupa
VinčaTY - JOUR AU - Rudolf, Rebeka AU - Stamenkovic, Dragoslav AU - Aleksić, Zoran AU - Jenko, Monika AU - Dordevic, Igor AU - Todorović, Aleksandar AU - Jokanović, Vukoman R. AU - Raic, Karlo T. PY - 2015 UR - https://vinar.vin.bg.ac.rs/handle/123456789/427 AB - Thin hydroxyapatite coatings were produced on Cp-Titanium Grade-2 samples, with new high-voltage pulse-power equipment PJ-100 (Plasma Jet, Serbia) in order to get a more stable implant structure appropriate for further clinical applications. A comparative analysis of differently prepared surfaces of the Cp-Titanium Grade-2 samples was done before the hydroxyapatite was applied. Microstructural observation of the modified hydroxyapatite/implant surface was done using scanning-electron-microscopy imaging and Auger electron spectroscopy, with the aim of detecting the morphology and the elements contained in the new surfaces of the samples. The results confirmed that the surface of Cp-Titanium Grade-2 modified with hydroxyapatite is very similar to the bone structure. T2 - Materials and Technology = Materiali in Tehnologije T1 - Hydroxyapatite Coatings on Cp-Titanium Grade-2 Surfaces Prepared with Plasma Spraying VL - 49 IS - 1 SP - 81 EP - 86 UR - https://hdl.handle.net/21.15107/rcub_vinar_427 ER -
@article{ author = "Rudolf, Rebeka and Stamenkovic, Dragoslav and Aleksić, Zoran and Jenko, Monika and Dordevic, Igor and Todorović, Aleksandar and Jokanović, Vukoman R. and Raic, Karlo T.", year = "2015", abstract = "Thin hydroxyapatite coatings were produced on Cp-Titanium Grade-2 samples, with new high-voltage pulse-power equipment PJ-100 (Plasma Jet, Serbia) in order to get a more stable implant structure appropriate for further clinical applications. A comparative analysis of differently prepared surfaces of the Cp-Titanium Grade-2 samples was done before the hydroxyapatite was applied. Microstructural observation of the modified hydroxyapatite/implant surface was done using scanning-electron-microscopy imaging and Auger electron spectroscopy, with the aim of detecting the morphology and the elements contained in the new surfaces of the samples. The results confirmed that the surface of Cp-Titanium Grade-2 modified with hydroxyapatite is very similar to the bone structure.", journal = "Materials and Technology = Materiali in Tehnologije", title = "Hydroxyapatite Coatings on Cp-Titanium Grade-2 Surfaces Prepared with Plasma Spraying", volume = "49", number = "1", pages = "81-86", url = "https://hdl.handle.net/21.15107/rcub_vinar_427" }
Rudolf, R., Stamenkovic, D., Aleksić, Z., Jenko, M., Dordevic, I., Todorović, A., Jokanović, V. R.,& Raic, K. T.. (2015). Hydroxyapatite Coatings on Cp-Titanium Grade-2 Surfaces Prepared with Plasma Spraying. in Materials and Technology = Materiali in Tehnologije, 49(1), 81-86. https://hdl.handle.net/21.15107/rcub_vinar_427
Rudolf R, Stamenkovic D, Aleksić Z, Jenko M, Dordevic I, Todorović A, Jokanović VR, Raic KT. Hydroxyapatite Coatings on Cp-Titanium Grade-2 Surfaces Prepared with Plasma Spraying. in Materials and Technology = Materiali in Tehnologije. 2015;49(1):81-86. https://hdl.handle.net/21.15107/rcub_vinar_427 .
Rudolf, Rebeka, Stamenkovic, Dragoslav, Aleksić, Zoran, Jenko, Monika, Dordevic, Igor, Todorović, Aleksandar, Jokanović, Vukoman R., Raic, Karlo T., "Hydroxyapatite Coatings on Cp-Titanium Grade-2 Surfaces Prepared with Plasma Spraying" in Materials and Technology = Materiali in Tehnologije, 49, no. 1 (2015):81-86, https://hdl.handle.net/21.15107/rcub_vinar_427 .