Application of TL Dosimeters for testing the accuracy of delivered dose in blood irradiator
Primena TL dozimetara za ispitivanje tačnosti isporučene doze u ozračivaču krvi
Конференцијски прилог (Објављена верзија)
Метаподаци
Приказ свих података о документуАпстракт
The aim of this work is to examine the accuracy of the delivered dose inside the blood irradiator using TL dosimeters. The TL dosimeters were placed behind the lead plate and were irradiated with absorbed dose values ranging from 50 mGy to 5 Gy in the 60Co reference radiation field. The deviations of the TL dosimeter readings in relation to the delivered values of the absorbed dose range from 7.8 % to 19.6 % with satisfactory linearity over the entire range of delivered doses (R=0.997). Monte Carlo simulations were used to determine the conversion coefficients for the application of this method in the 137Cs radiation field for devices that work with this type of radiation source.
Cilj ovog rada je ispitivanje tačnosti isporučene doze unutar ozračivača krvi koristeći TL dozimetre. TL dozimetri su postavljeni iza olovne ploče i ozračeni su vrednostima apsorbovane doze u opsegu od 50 mGy do 5 Gy u referentnom polju zračenja 60Co. Odstupanja očitavanja TL dozimetara u odnosu na isporučene vrednosti apsorbovane doze su u opsegu od 7.8% do 19.6% uz zadovoljavajuću linearnost na celom opsegu isporučenih doza (R=0.997). Monte Karlo simulacijama su određeni konverzioni koeficijenti za primenu ovog metoda i u polju zračenja 137Cs za uređaje koji rade sa ovom vrstom izvora zračenja.
Извор:
32. simpozijum DZZSCG : zbornik radova, 2023, 387-393Издавач:
- Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе
Напомена:
- XXXII симпозијум ДЗЗСЦГ (Друштва за заштиту од зрачења Србије и Црне Горе), 4-6. октобар 2023. године
Колекције
Институција/група
VinčaTY - CONF AU - Božović, Predrag AU - Kojić, Andrea AU - Kržanović, Nikola AU - Topalović, Dušan AU - Stanković Petrović, Jelena PY - 2023 UR - https://vinar.vin.bg.ac.rs/handle/123456789/11928 AB - The aim of this work is to examine the accuracy of the delivered dose inside the blood irradiator using TL dosimeters. The TL dosimeters were placed behind the lead plate and were irradiated with absorbed dose values ranging from 50 mGy to 5 Gy in the 60Co reference radiation field. The deviations of the TL dosimeter readings in relation to the delivered values of the absorbed dose range from 7.8 % to 19.6 % with satisfactory linearity over the entire range of delivered doses (R=0.997). Monte Carlo simulations were used to determine the conversion coefficients for the application of this method in the 137Cs radiation field for devices that work with this type of radiation source. AB - Cilj ovog rada je ispitivanje tačnosti isporučene doze unutar ozračivača krvi koristeći TL dozimetre. TL dozimetri su postavljeni iza olovne ploče i ozračeni su vrednostima apsorbovane doze u opsegu od 50 mGy do 5 Gy u referentnom polju zračenja 60Co. Odstupanja očitavanja TL dozimetara u odnosu na isporučene vrednosti apsorbovane doze su u opsegu od 7.8% do 19.6% uz zadovoljavajuću linearnost na celom opsegu isporučenih doza (R=0.997). Monte Karlo simulacijama su određeni konverzioni koeficijenti za primenu ovog metoda i u polju zračenja 137Cs za uređaje koji rade sa ovom vrstom izvora zračenja. PB - Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе C3 - 32. simpozijum DZZSCG : zbornik radova T1 - Application of TL Dosimeters for testing the accuracy of delivered dose in blood irradiator T1 - Primena TL dozimetara za ispitivanje tačnosti isporučene doze u ozračivaču krvi SP - 387 EP - 393 UR - https://hdl.handle.net/21.15107/rcub_vinar_11928 ER -
@conference{ author = "Božović, Predrag and Kojić, Andrea and Kržanović, Nikola and Topalović, Dušan and Stanković Petrović, Jelena", year = "2023", abstract = "The aim of this work is to examine the accuracy of the delivered dose inside the blood irradiator using TL dosimeters. The TL dosimeters were placed behind the lead plate and were irradiated with absorbed dose values ranging from 50 mGy to 5 Gy in the 60Co reference radiation field. The deviations of the TL dosimeter readings in relation to the delivered values of the absorbed dose range from 7.8 % to 19.6 % with satisfactory linearity over the entire range of delivered doses (R=0.997). Monte Carlo simulations were used to determine the conversion coefficients for the application of this method in the 137Cs radiation field for devices that work with this type of radiation source., Cilj ovog rada je ispitivanje tačnosti isporučene doze unutar ozračivača krvi koristeći TL dozimetre. TL dozimetri su postavljeni iza olovne ploče i ozračeni su vrednostima apsorbovane doze u opsegu od 50 mGy do 5 Gy u referentnom polju zračenja 60Co. Odstupanja očitavanja TL dozimetara u odnosu na isporučene vrednosti apsorbovane doze su u opsegu od 7.8% do 19.6% uz zadovoljavajuću linearnost na celom opsegu isporučenih doza (R=0.997). Monte Karlo simulacijama su određeni konverzioni koeficijenti za primenu ovog metoda i u polju zračenja 137Cs za uređaje koji rade sa ovom vrstom izvora zračenja.", publisher = "Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе", journal = "32. simpozijum DZZSCG : zbornik radova", title = "Application of TL Dosimeters for testing the accuracy of delivered dose in blood irradiator, Primena TL dozimetara za ispitivanje tačnosti isporučene doze u ozračivaču krvi", pages = "387-393", url = "https://hdl.handle.net/21.15107/rcub_vinar_11928" }
Božović, P., Kojić, A., Kržanović, N., Topalović, D.,& Stanković Petrović, J.. (2023). Application of TL Dosimeters for testing the accuracy of delivered dose in blood irradiator. in 32. simpozijum DZZSCG : zbornik radova Београд : Институт за нуклеарне науке "Винча" : Друштво за заштиту од зрачења Србије и Црне Горе., 387-393. https://hdl.handle.net/21.15107/rcub_vinar_11928
Božović P, Kojić A, Kržanović N, Topalović D, Stanković Petrović J. Application of TL Dosimeters for testing the accuracy of delivered dose in blood irradiator. in 32. simpozijum DZZSCG : zbornik radova. 2023;:387-393. https://hdl.handle.net/21.15107/rcub_vinar_11928 .
Božović, Predrag, Kojić, Andrea, Kržanović, Nikola, Topalović, Dušan, Stanković Petrović, Jelena, "Application of TL Dosimeters for testing the accuracy of delivered dose in blood irradiator" in 32. simpozijum DZZSCG : zbornik radova (2023):387-393, https://hdl.handle.net/21.15107/rcub_vinar_11928 .