Harmonization of IEC type testing requirements and test methods for active area dosimeters in environmental monitoring
Само за регистроване кориснике
2022
Аутори
Kržanović, NikolaZutz, Hayo
Hupe, Oliver
Živanović, Miloš Z.
Ketelhut, Steffen
Luchkov, Maksym
Чланак у часопису (Објављена верзија)
Метаподаци
Приказ свих података о документуАпстракт
Advancements in area monitoring caused by development of environmental monitoring networks which utilise active electronic dosimeters have led to widespread use of active area dosimeters. Currently, there are various sets of requirements for area dosimeters available in IEC standards. A clear distinction is not always made between requirements for area dosimeters used for workplace monitoring and that used for environmental monitoring. Performance testing methods related to the radiation characteristics were examined in a number of standards. Out of this, criteria in terms of relative response limits of variation have been derived for photon energy, angle of incidence and dose rate non-linearity for both dosimeter use cases. Testing of several active area dosimeters which are used for environmental monitoring has been performed in a wide range of photon energy, angle of incidence and dose rate. The collected requirements for environmental and workplace area dosimetry imply that some ha...rmonization of IEC standards for area dosimeters is needed.
Кључне речи:
Area monitoring / Environmental monitoring / Harmonization / Performance testing / Radiation protection / StandardizationИзвор:
Radiation Physics and Chemistry, 2022, 198, 110291-Финансирање / пројекти:
- RMG - DOSEtrace project [17RPT01]
DOI: 10.1016/j.radphyschem.2022.110291
ISSN: 0969-806X
WoS: 00083353930000
Scopus: 2-s2.0-85131698040
Колекције
Институција/група
VinčaTY - JOUR AU - Kržanović, Nikola AU - Zutz, Hayo AU - Hupe, Oliver AU - Živanović, Miloš Z. AU - Ketelhut, Steffen AU - Luchkov, Maksym PY - 2022 UR - https://vinar.vin.bg.ac.rs/handle/123456789/10298 AB - Advancements in area monitoring caused by development of environmental monitoring networks which utilise active electronic dosimeters have led to widespread use of active area dosimeters. Currently, there are various sets of requirements for area dosimeters available in IEC standards. A clear distinction is not always made between requirements for area dosimeters used for workplace monitoring and that used for environmental monitoring. Performance testing methods related to the radiation characteristics were examined in a number of standards. Out of this, criteria in terms of relative response limits of variation have been derived for photon energy, angle of incidence and dose rate non-linearity for both dosimeter use cases. Testing of several active area dosimeters which are used for environmental monitoring has been performed in a wide range of photon energy, angle of incidence and dose rate. The collected requirements for environmental and workplace area dosimetry imply that some harmonization of IEC standards for area dosimeters is needed. T2 - Radiation Physics and Chemistry T1 - Harmonization of IEC type testing requirements and test methods for active area dosimeters in environmental monitoring VL - 198 SP - 110291 DO - 10.1016/j.radphyschem.2022.110291 ER -
@article{ author = "Kržanović, Nikola and Zutz, Hayo and Hupe, Oliver and Živanović, Miloš Z. and Ketelhut, Steffen and Luchkov, Maksym", year = "2022", abstract = "Advancements in area monitoring caused by development of environmental monitoring networks which utilise active electronic dosimeters have led to widespread use of active area dosimeters. Currently, there are various sets of requirements for area dosimeters available in IEC standards. A clear distinction is not always made between requirements for area dosimeters used for workplace monitoring and that used for environmental monitoring. Performance testing methods related to the radiation characteristics were examined in a number of standards. Out of this, criteria in terms of relative response limits of variation have been derived for photon energy, angle of incidence and dose rate non-linearity for both dosimeter use cases. Testing of several active area dosimeters which are used for environmental monitoring has been performed in a wide range of photon energy, angle of incidence and dose rate. The collected requirements for environmental and workplace area dosimetry imply that some harmonization of IEC standards for area dosimeters is needed.", journal = "Radiation Physics and Chemistry", title = "Harmonization of IEC type testing requirements and test methods for active area dosimeters in environmental monitoring", volume = "198", pages = "110291", doi = "10.1016/j.radphyschem.2022.110291" }
Kržanović, N., Zutz, H., Hupe, O., Živanović, M. Z., Ketelhut, S.,& Luchkov, M.. (2022). Harmonization of IEC type testing requirements and test methods for active area dosimeters in environmental monitoring. in Radiation Physics and Chemistry, 198, 110291. https://doi.org/10.1016/j.radphyschem.2022.110291
Kržanović N, Zutz H, Hupe O, Živanović MZ, Ketelhut S, Luchkov M. Harmonization of IEC type testing requirements and test methods for active area dosimeters in environmental monitoring. in Radiation Physics and Chemistry. 2022;198:110291. doi:10.1016/j.radphyschem.2022.110291 .
Kržanović, Nikola, Zutz, Hayo, Hupe, Oliver, Živanović, Miloš Z., Ketelhut, Steffen, Luchkov, Maksym, "Harmonization of IEC type testing requirements and test methods for active area dosimeters in environmental monitoring" in Radiation Physics and Chemistry, 198 (2022):110291, https://doi.org/10.1016/j.radphyschem.2022.110291 . .